Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model
Pediatric Pulmonology · 2025-04 · 60 (4)
DOI: 10.1002/ppul.71090 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Pediatric Pulmonology, online erschienen am 11.04.2025. Verfasst von F-X Anzinger und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatric Surgery, Ludwig-Maximilians-University of Munich | LMU, Munich, Germany. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Chest Tubes, Drainage, Models, Anatomic. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
In recent years, digital chest tube drainage systems have been introduced. Limited studies address their benefits and risks in pediatric patients, particularly neonates. This study compares a three‐chamber chest drainage system with a digital system using a high‐fidelity 3D‐printed model.
Methods
We conducted direct measurements and 3D‐printed model tests with both systems at different suction pressures (−1 to −20 cmH 2 O) to assess the actual pressures. The effects of siphon and automatic flushes in the digital system were also studied.
Results
At −20 and −10 cmH 2 O, significant differences were found between the digital and analog systems in direct and model measurements. The analog system became unreliable below −10 cmH 2 O. For the digital system, most measurements remained within the set pressures, with outliers up to −30 cmH 2 O due to regular flushing.
Conclusion
This experimental study evaluates the suitability of digital drainage systems for the pediatric and neonatal populations. Our model demonstrated reliable simulation of thoracic conditions, making it a useful tool for pre‐clinical testing where patient testing may be limited. Both systems yielded satisfactory results at −20 and −10 cmH 2 O, but the digital system showed greater flexibility, maintaining pressures as low as −5 cmH 2 O. The analog system was consistent but less adaptable, which may limit its use in dynamic situations. The digital system's ability to simulate more flexible scenarios offers potential clinical advantages, though further investigation is needed to assess its impact on neonatal safety. The increase in suction during flushing may pose a risk for neonatal patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pediatric Surgery, Ludwig-Maximilians-University of Munich | LMU, Munich, Germany.
- Ludwig-Maximilians-University of Munich | LMU, Chair for Experimental Physics, Munich, Germany.
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Anzinger, F. X., Hashagen, T. J., Palaniappan, P., Lindner, A., Riboldi, M., Gödeke, J., & Muensterer, O. J. (2025). Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model. Pediatric Pulmonology, 60(4). https://doi.org/10.1002/ppul.71090
Vancouver
Anzinger FX, Hashagen TJ, Palaniappan P, Lindner A, Riboldi M, Gödeke J, et al. Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model. Pediatric Pulmonology. 2025;60(4). doi:10.1002/ppul.71090
BibTeX
@article{Anzinger2025,
author = {Anzinger, F-X and Hashagen, T J and Palaniappan, P and Lindner, A and Riboldi, M and Gödeke, J and Muensterer, O J},
title = {Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model},
journal = {Pediatric Pulmonology},
year = {2025},
volume = {60},
number = {4},
doi = {10.1002/ppul.71090}
}