EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Preclinical Efficacy and Safety Evaluation of a Novel Controlled‐Expansion ePTFE‐Covered TIPS Stent in a Swine Model

Yi Xiang, Qi Zhang, Chuan Liu, Jiawei Zhang, Qingqiao Zhang, Changsheng Shi, Chunze Zhou, Yudong Zhang, Hongliang He, Hao Xie, Hui Shi, Haijian Guo und 4 weitere

Portal Hypertension & Cirrhosis · 15.05.2026

DOI: 10.1002/poh2.70052 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Portal Hypertension & Cirrhosis, online erschienen am 15.05.2026. Verfasst von Yi Xiang und 15 weiteren Autor:innen. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Jiangsu Province.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Conventional fixed‐caliber transjugular intrahepatic portosystemic shunt (TIPS) stents may limit individualized portal decompression and lead to over‐ or under‐shunting. We therefore aimed to evaluate the safety and performance of a novel controlled‐expansion expanded polytetrafluoroethylene (ePTFE)‐covered TIPS stent, which is designed for precise placement, secure anchoring, and on‐demand diameter adjustment.

Methods

Eleven healthy swine (40–50 kg) underwent TIPS with a self‐expanding nitinol stent‐graft comprising a covered intrahepatic segment and an uncovered portal segment, with balloon‐calibrated diameter adjustment across 6.0–10.0 mm; all stents were post‐dilated to a fixed 8.0‐mm functional diameter. Animals were assigned to 1‐ ( n = 3), 3‐ ( n = 3), or 6‐month ( n = 5) endpoints. Patency was defined as < 50% diameter stenosis without angiographic occlusion. Patency and integrity were assessed by digital subtraction angiography with quantitative vascular angiography (QVA). Systemic and local responses were evaluated by clinical monitoring, hematology/serum biochemistry, comprehensive necropsy, and blinded histopathology.

Results

Technical success rate was 100% in all 11 cases, with no intraoperative complications; all animals survived to endpoint. Deployment, visibility, and handling were consistently excellent. Follow‐up angiography showed patent shunts without migration, thrombosis, or fracture; no collapse, kinking, or ePTFE delamination. QVA demonstrated mild luminal loss: mean diameter stenosis was 4.88 (3.13–7.19)% at 1 month, 9.93 (7.30–14.09)% at 3 months, and 13.41 (12.05–16.05)% at 6 months; all shunts showed < 30% stenosis. Laboratory values remained within or returned to normal limits. Necropsy confirmed well‐incorporated tracts without device‐related pathology. Histology showed complete endothelialization, a thin neointima with minimal inflammation, and absence of thrombosis at all time points.

Conclusions

In healthy swine, this stent demonstrated favorable safety and patency; performance in portal hypertension remains to be established. This adjustable‐caliber design may help overcome some limitations of fixed‐diameter stents and supports further evaluation in disease models and humans.

Förderung

  • National Natural Science Foundation of China
  • Natural Science Foundation of Jiangsu Province

Zitieren

APA

Xiang, Y., Zhang, Q., Liu, C., Zhang, J., Zhang, Q., Shi, C., Zhou, C., Zhang, Y., He, H., Xie, H., Shi, H., Guo, H., Liu, W., Lin, M., Yu, H., & Qi, X. (2026). Preclinical Efficacy and Safety Evaluation of a Novel Controlled‐Expansion ePTFE‐Covered TIPS Stent in a Swine Model. Portal Hypertension & Cirrhosis. https://doi.org/10.1002/poh2.70052

Vancouver

Xiang Y, Zhang Q, Liu C, Zhang J, Zhang Q, Shi C, et al. Preclinical Efficacy and Safety Evaluation of a Novel Controlled‐Expansion ePTFE‐Covered TIPS Stent in a Swine Model. Portal Hypertension & Cirrhosis. 2026. doi:10.1002/poh2.70052

BibTeX
@article{Xiang2026,
  author = {Xiang, Yi and Zhang, Qi and Liu, Chuan and Zhang, Jiawei and Zhang, Qingqiao and Shi, Changsheng and Zhou, Chunze and Zhang, Yudong and He, Hongliang and Xie, Hao and Shi, Hui and Guo, Haijian and Liu, Wenjing and Lin, Muqing and Yu, Huajie and Qi, Xiaolong},
  title = {Preclinical Efficacy and Safety Evaluation of a Novel Controlled‐Expansion ePTFE‐Covered TIPS Stent in a Swine Model},
  journal = {Portal Hypertension & Cirrhosis},
  year = {2026},
  doi = {10.1002/poh2.70052}
}
Kostenlos starten →