Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall

Kalena Liu iD, Lauren O'Brien, Dan Campbell, Rebecca Chiffer, Thomas Willcox, Jacob B Hunter

Otolaryngology–Head and Neck Surgery · 2026

DOI: 10.1002/ohn.70048 ↗

At a glance

Journal article in Otolaryngology–Head and Neck Surgery, 2026. Written by Kalena Liu and 5 co-authors. Institution: Department of Otolaryngology-Head and Neck Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania, USA. Major MEDLINE topics: Cochlear Implants, Prosthesis Failure, Cochlear Implantation, Medical Device Recalls. 13 references.

PovzetekSubjectsBibliografski podatkiCiteRelated studies

Abstract (original language)

Objective

To characterize the timing and likelihood of electrode failure in Advanced Bionics (AB) cochlear implants and assess long‐term trends following the 2020 recall of HiRes Ultra and Ultra 3D models. Study Design Retrospective cohort study.

Setting

Single tertiary academic medical center.

Methods

We conducted a retrospective chart review of patients who received AB HiRes Ultra or Ultra 3D cochlear AB implants between January 2014 and February 2020. Patients who experienced device failure requiring revision surgery were identified. Demographics, implant date, failure status, and time to failure were recorded. Kaplan‐Meier survival analysis was used to model cumulative failure risk over time.

Results

A total of 140 AB implants were placed, with 24 (17.1%) device failures identified. The median time to failure was 3.6 years (range: 0.39‐6.0 years). Median follow‐up was 5.2 years (range: 0.1‐10.7 years). There was no significant association between failure and gender, age, race/ethnicity, or laterality. Kaplan‐Meier analysis demonstrated that 95% of failures occurred within the first 5.9 years, and the failure rate decreased to 2.1% per year after 5 years.

Conclusion

AB cochlear implant failures primarily occur within the first 5 years postimplantation, with significantly reduced risk beyond this period. These findings support long‐term device stability and can inform patient counseling, expectations, and follow‐up protocols.

Subjects

MeSH terms (MEDLINE)

Cochlear ImplantsProsthesis FailureCochlear ImplantationMedical Device RecallsHumansRetrospective StudiesMaleFemaleMiddle AgedAdultAgedAdolescentChildChild, PreschoolReoperationTime FactorsYoung AdultProsthesis DesignInfant

Author keywords

cochlear implantsdevice recallelectrode failure

Authors and institutions

  • Department of Otolaryngology-Head and Neck Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.

ORCID profiles: Kalena Liu

Cite

APA

Liu, K., O'Brien, L., Campbell, D., Chiffer, R., Willcox, T., & Hunter, J. B. (2026). Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.70048

Vancouver

Liu K, O'Brien L, Campbell D, Chiffer R, Willcox T, Hunter JB. Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall. Otolaryngology–Head and Neck Surgery. 2026. doi:10.1002/ohn.70048

BibTeX
@article{Liu2026,
  author = {Liu, Kalena and O'Brien, Lauren and Campbell, Dan and Chiffer, Rebecca and Willcox, Thomas and Hunter, Jacob B},
  title = {Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall},
  journal = {Otolaryngology–Head and Neck Surgery},
  year = {2026},
  doi = {10.1002/ohn.70048}
}