Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall
Otolaryngology–Head and Neck Surgery · 2026
DOI: 10.1002/ohn.70048 ↗Kurz gefasst
Fachartikel in der Zeitschrift Otolaryngology–Head and Neck Surgery, 2026. Verfasst von Kalena Liu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Cochlear Implants, Prosthesis Failure, Cochlear Implantation, Medical Device Recalls. Die Arbeit nennt 13 Literaturangaben.
Abstract (Originalsprache)
Objective
To characterize the timing and likelihood of electrode failure in Advanced Bionics (AB) cochlear implants and assess long‐term trends following the 2020 recall of HiRes Ultra and Ultra 3D models. Study Design Retrospective cohort study.
Setting
Single tertiary academic medical center.
Methods
We conducted a retrospective chart review of patients who received AB HiRes Ultra or Ultra 3D cochlear AB implants between January 2014 and February 2020. Patients who experienced device failure requiring revision surgery were identified. Demographics, implant date, failure status, and time to failure were recorded. Kaplan‐Meier survival analysis was used to model cumulative failure risk over time.
Results
A total of 140 AB implants were placed, with 24 (17.1%) device failures identified. The median time to failure was 3.6 years (range: 0.39‐6.0 years). Median follow‐up was 5.2 years (range: 0.1‐10.7 years). There was no significant association between failure and gender, age, race/ethnicity, or laterality. Kaplan‐Meier analysis demonstrated that 95% of failures occurred within the first 5.9 years, and the failure rate decreased to 2.1% per year after 5 years.
Conclusion
AB cochlear implant failures primarily occur within the first 5 years postimplantation, with significantly reduced risk beyond this period. These findings support long‐term device stability and can inform patient counseling, expectations, and follow‐up protocols.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
ORCID-Profile: Kalena Liu
Zitieren
APA
Liu, K., O'Brien, L., Campbell, D., Chiffer, R., Willcox, T., & Hunter, J. B. (2026). Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.70048
Vancouver
Liu K, O'Brien L, Campbell D, Chiffer R, Willcox T, Hunter JB. Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall. Otolaryngology–Head and Neck Surgery. 2026. doi:10.1002/ohn.70048
BibTeX
@article{Liu2026,
author = {Liu, Kalena and O'Brien, Lauren and Campbell, Dan and Chiffer, Rebecca and Willcox, Thomas and Hunter, Jacob B},
title = {Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2026},
doi = {10.1002/ohn.70048}
}
EUVIMED