Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies

Paolo Piaggi iD, Theresa L Rodzevik, Erica Wohlers, Katherine Ruud, Jon Moon, Jonathan Krakoff, Douglas C Chang

Obesity · 2023

DOI: 10.1002/oby.23683 ↗

At a glance

Journal article in Obesity, 2023. Written by Paolo Piaggi and 6 co-authors. Institution: Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch, National Institute of Diabetes and Digestive and K…. Major MEDLINE topics: Carbon Dioxide, Oxygen. Cited 2 times so far (Crossref); 21 references. Funded by: National Institute of Diabetes and Digestive and Kidney Diseases, National Institutes of Health.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Objective

The aim of this study was to test proportional‐integral‐derivative (PID) control of air inflow rate in a whole‐room indirect calorimeter to improve accuracy in measuring oxygen (O 2 ) consumption () and carbon dioxide (CO 2 ) production ().

Methods

A precision gas blender infused nitrogen (N 2 ) and CO 2 into the calorimeter over 24 hours based on static and dynamic infusion profiles mimicking and patterns during resting and non‐resting conditions. Constant (60 L/min) versus time‐variant flow set by a PID controller based on the CO 2 concentration was compared based on errors between measured versus expected values for respiratory exchange ratio, and metabolic rate.

Results

Compared with constant inflow, the PID controller allowed both a faster rise time and long‐term maintenance of a stable CO 2 concentration inside the calorimeter, resulting in more accurate estimates (mean hourly error, PID: −0.9%, 60 L/min = −2.3%, p < 0.05) during static infusions. During dynamic infusions mimicking exercise sessions, the PID controller achieved smaller errors for (mean: −0.6% vs. −2.7%, p = 0.02) and respiratory exchange ratio (mean: 0.5% vs. −3.1%, p = 0.02) compared with constant inflow conditions, with similar ( p = 0.97) and metabolic rate ( p = 0.76) errors.

Conclusions

PID control in a whole‐room indirect calorimeter system leads to more accurate measurements of substrate oxidation during dynamic metabolic studies.

Subjects

MeSH terms (MEDLINE)

Carbon DioxideOxygenEnergy MetabolismOxygen ConsumptionTime FactorsCalorimetry, Indirect

Substances

Carbon DioxideOxygen

Authors and institutions

  • Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch, National Institute of Diabetes and Digestive and Kidney Diseases, Phoenix, Arizona, USA.
  • Department of Information Engineering, University of Pisa, Pisa, Italy.
  • MEI Research, Ltd., Edina, Minnesota, USA.

ORCID profiles: Paolo Piaggi

Funding

  • National Institute of Diabetes and Digestive and Kidney Diseases
  • National Institutes of Health

Cite

APA

Piaggi, P., Rodzevik, T. L., Wohlers, E., Ruud, K., Moon, J., Krakoff, J., & Chang, D. C. (2023). Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies. Obesity. https://doi.org/10.1002/oby.23683

Vancouver

Piaggi P, Rodzevik TL, Wohlers E, Ruud K, Moon J, Krakoff J, et al. Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies. Obesity. 2023. doi:10.1002/oby.23683

BibTeX
@article{Piaggi2023,
  author = {Piaggi, Paolo and Rodzevik, Theresa L and Wohlers, Erica and Ruud, Katherine and Moon, Jon and Krakoff, Jonathan and Chang, Douglas C},
  title = {Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies},
  journal = {Obesity},
  year = {2023},
  doi = {10.1002/oby.23683}
}
Start for free →