Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies
Obesity · 2023
DOI: 10.1002/oby.23683 ↗Kurz gefasst
Fachartikel in der Zeitschrift Obesity, 2023. Verfasst von Paolo Piaggi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch, National Institute of Diabetes and Digestive and K…. Hauptthemen laut MEDLINE: Carbon Dioxide, Oxygen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Gefördert von: National Institute of Diabetes and Digestive and Kidney Diseases, National Institutes of Health.
Abstract (Originalsprache)
Objective
The aim of this study was to test proportional‐integral‐derivative (PID) control of air inflow rate in a whole‐room indirect calorimeter to improve accuracy in measuring oxygen (O 2 ) consumption () and carbon dioxide (CO 2 ) production ().
Methods
A precision gas blender infused nitrogen (N 2 ) and CO 2 into the calorimeter over 24 hours based on static and dynamic infusion profiles mimicking and patterns during resting and non‐resting conditions. Constant (60 L/min) versus time‐variant flow set by a PID controller based on the CO 2 concentration was compared based on errors between measured versus expected values for respiratory exchange ratio, and metabolic rate.
Results
Compared with constant inflow, the PID controller allowed both a faster rise time and long‐term maintenance of a stable CO 2 concentration inside the calorimeter, resulting in more accurate estimates (mean hourly error, PID: −0.9%, 60 L/min = −2.3%, p < 0.05) during static infusions. During dynamic infusions mimicking exercise sessions, the PID controller achieved smaller errors for (mean: −0.6% vs. −2.7%, p = 0.02) and respiratory exchange ratio (mean: 0.5% vs. −3.1%, p = 0.02) compared with constant inflow conditions, with similar ( p = 0.97) and metabolic rate ( p = 0.76) errors.
Conclusions
PID control in a whole‐room indirect calorimeter system leads to more accurate measurements of substrate oxidation during dynamic metabolic studies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Obesity and Diabetes Clinical Research Section, Phoenix Epidemiology and Clinical Research Branch, National Institute of Diabetes and Digestive and Kidney Diseases, Phoenix, Arizona, USA.
- Department of Information Engineering, University of Pisa, Pisa, Italy.
- MEI Research, Ltd., Edina, Minnesota, USA.
ORCID-Profile: Paolo Piaggi
Förderung
- National Institute of Diabetes and Digestive and Kidney Diseases
- National Institutes of Health
Zitieren
APA
Piaggi, P., Rodzevik, T. L., Wohlers, E., Ruud, K., Moon, J., Krakoff, J., & Chang, D. C. (2023). Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies. Obesity. https://doi.org/10.1002/oby.23683
Vancouver
Piaggi P, Rodzevik TL, Wohlers E, Ruud K, Moon J, Krakoff J, et al. Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies. Obesity. 2023. doi:10.1002/oby.23683
BibTeX
@article{Piaggi2023,
author = {Piaggi, Paolo and Rodzevik, Theresa L and Wohlers, Erica and Ruud, Katherine and Moon, Jon and Krakoff, Jonathan and Chang, Douglas C},
title = {Closed‐loop control of air supply to whole‐room indirect calorimeters to improve accuracy and standardize measurements during 24‐hour dynamic metabolic studies},
journal = {Obesity},
year = {2023},
doi = {10.1002/oby.23683}
}
EUVIMED