Fachartikel in der Zeitschrift NMR in Biomedicine, online erschienen am 03.08.2026. Verfasst von Zakaria Zariry und 10 weiteren Autor:innen. Beteiligte Einrichtung: CNRS, Institute of Cognitive Sciences Marc Jeannerod, Bron, France. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Head Movements, Head. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NICHD NIH HHS, LABEX CORTEX, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Despite numerous magnetic resonance imaging (MRI) head motion mitigation strategies, the lack of rigorous evaluation limits their optimization and clinical adoption. We propose an in vivo framework combining a visual instruction system for reproducible head motion with reference standard interpose displacement estimation to assess intra‐MRI tracking accuracy and precision. Its utility is demonstrated by comparing a markerless optical system (MOS) and a fat‐signal navigator (FatNav). Six participants underwent 3T T1‐weighted brain MRI with a FatNav module, performing visually guided 2° and 4° head rotations around the X ‐ and Z ‐axes using MOS feedback. T1‐weighted images were acquired at seven distinct head poses. MOS and FatNav motion estimates were compared against rigid registration of the T1‐weighted images, which served as the reference standard. MOS‐ and FatNav‐corrected images for the three successive head rotations were also compared using the structural similarity index measure (SSIM), peak signal‐to‐noise ratio (PSNR), and a focus measure. FatNav accuracy was inferior for translations ( p < 0.001) and 2°–4° rotations but improved to match MOS for subtle pitch + and yaw + , even surpassing it for subtle yaw − . Meanwhile, MOS precision was higher for yaw + than yaw − ( p < 0.001) but inferior to FatNav for pitch + ( p = 0.041). MOS better restored T1‐weighted image fidelity, yielding higher SSIM, PSNR, and focus ( p < 0.01). Notably, the framework detected a subtle improvement in FatNav performance with neck masking, an effect uncaptured by conventional image quality metrics. In conclusion, while image quality metrics suggested superior overall correction with MOS, our framework provided a more detailed characterization of in vivo performance differences.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingHead MovementsHeadHumansMaleSignal-To-Noise RatioAdultMotion CaptureReproducibility of ResultsFemaleArtifactsBrainMotion

Stichwörter der Autor:innen

brain MRIfat navigatormarkerless optical systemmotion artifactsmotion correctionmovement tracking

Autor:innen und Institutionen

  • CNRS, Institute of Cognitive Sciences Marc Jeannerod, Bron, France.
  • Université Claude Bernard, Lyon 1, Villeurbanne, France.
  • Cermep, CNRS/UAR 3453, Lyon, France.
  • SFR Santé Lyon-Est, CNRS UAR3453, INSERM US7, Université Lyon 1, Lyon, France.
  • Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.

ORCID-Profile: Zakaria Zariry, Franck Lamberton, Robert Frost, Thomas Troalen, Jakob M Slipsager, Andre van der Kouwe, James Bonaiuto, Bassem Hiba

Förderung

  • NICHD NIH HHS
  • LABEX CORTEX
  • NIBIB NIH HHS
  • European Research Council
  • Eunice Kennedy Shriver National Institute of Child Health and Human Development
  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Zariry, Z., Lamberton, F., Frost, R., Gaass, T., Troalen, T., Rayson, H., Slipsager, J. M., Richard, N., van der Kouwe, A., Bonaiuto, J., & Hiba, B. (2026). Intra‐MRI Head Motion Tracking and Correction: A Quantitative In Vivo Evaluation Framework. NMR in Biomedicine, 39(9). https://doi.org/10.1002/nbm.70368

Vancouver

Zariry Z, Lamberton F, Frost R, Gaass T, Troalen T, Rayson H, et al. Intra‐MRI Head Motion Tracking and Correction: A Quantitative In Vivo Evaluation Framework. NMR in Biomedicine. 2026;39(9). doi:10.1002/nbm.70368

BibTeX
@article{Zariry2026,
  author = {Zariry, Zakaria and Lamberton, Franck and Frost, Robert and Gaass, Thomas and Troalen, Thomas and Rayson, Holly and Slipsager, Jakob M and Richard, Nathalie and van der Kouwe, Andre and Bonaiuto, James and Hiba, Bassem},
  title = {Intra‐MRI Head Motion Tracking and Correction: A Quantitative In Vivo Evaluation Framework},
  journal = {NMR in Biomedicine},
  year = {2026},
  volume = {39},
  number = {9},
  doi = {10.1002/nbm.70368}
}
Kostenlos starten →