Robust Magnetic Field Homogeneity Enhancement via Local Passive Shimming in a Superconducting Magnet for Animal MRI
NMR in Biomedicine · 2026-06 · 39 (6)
DOI: 10.1002/nbm.70298 ↗Kurz gefasst
Fachartikel in der Zeitschrift NMR in Biomedicine, online erschienen am 22.04.2026. Verfasst von Haoran Chen und 11 weiteren Autor:innen. Beteiligte Einrichtung: School of Electrical and Electronic Engineering, Huazhong University of Science and Technology, Wuhan, China. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Magnetic Fields, Superconductivity, Magnets. Die Arbeit nennt 29 Literaturangaben. Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
In ultrahigh‐field magnetic resonance imaging (MRI) systems, achieving subppm static magnetic field homogeneity is critical. However, engineering tolerances and assembly errors often introduce complex field inhomogeneities that global shimming alone cannot always fully correct. In this study, a novel local passive shimming method is developed for small‐bore MRI superconducting magnets. By establishing a linear programming model for local shimming, the approach enables region‐specific correction of high‐order field variations. Numerical simulations and experimental verification on a 7 T/300‐mm superconducting magnet demonstrate significant improvements in homogeneity, reducing root mean square error (RMSE) and peak‐to‐peak variation across the diameter of spherical volume (DSV). The optimized shim configuration achieved a peak‐to‐peak inhomogeneity of 4.23 ppm within the DSV, indicating substantial enhancement over baseline performance. These results indicate that the proposed technique serves as a practical and efficient complement to global shimming strategies, offering a robust solution for field correction in small‐bore MRI systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Electrical and Electronic Engineering, Huazhong University of Science and Technology, Wuhan, China.
- Beijing Key Laboratory of Applied Superconductivity, Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, China.
- School of Electronic, Electrical and Communication Engineering, University of Chinese Academy of Sciences, Beijing, China.
- School of Electrical Engineering and Computer Science, The University of Queensland, Brisbane, QLD, Australia.
ORCID-Profile: Haoran Chen, Yaohui Wang
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Chen, H., Wang, Y., Wang, W., Wang, H., Yang, W., Wei, S., Xu, C., Chen, S., Liu, J., Sun, J., Li, L., & Liu, F. (2026). Robust Magnetic Field Homogeneity Enhancement via Local Passive Shimming in a Superconducting Magnet for Animal MRI. NMR in Biomedicine, 39(6). https://doi.org/10.1002/nbm.70298
Vancouver
Chen H, Wang Y, Wang W, Wang H, Yang W, Wei S, et al. Robust Magnetic Field Homogeneity Enhancement via Local Passive Shimming in a Superconducting Magnet for Animal MRI. NMR in Biomedicine. 2026;39(6). doi:10.1002/nbm.70298
BibTeX
@article{Chen2026,
author = {Chen, Haoran and Wang, Yaohui and Wang, Wenchen and Wang, Hui and Yang, Wenhui and Wei, Shufeng and Xu, Ce and Chen, Shunzhong and Liu, Jianhua and Sun, Jinshui and Li, Liang and Liu, Feng},
title = {Robust Magnetic Field Homogeneity Enhancement via Local Passive Shimming in a Superconducting Magnet for Animal MRI},
journal = {NMR in Biomedicine},
year = {2026},
volume = {39},
number = {6},
doi = {10.1002/nbm.70298}
}
EUVIMED