Harmonic analysis and optimization for closed‐loop superconducting shim coils of 7 T MRI magnet
Medical Physics · 2025
DOI: 10.1002/mp.17641 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Zijie Lin und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Plasma Physics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, China. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Superconductivity. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Institute of Energy, Hefei Comprehensive National Science Center, Collaborative Innovation Project of Colleges and Universities of Anhui Province.
Abstract (Originalsprache)
Background
High‐resolution brain imaging is crucial in clinical diagnosis and neuroscience, with ultra‐high field strength MRI systems () offering significant advantages for imaging neuronal microstructures. However, achieving magnetic field homogeneity is challenging due to engineering faults during the installation of superconducting strip windings and the primary magnet.
Purpose
This study aims to design and optimize active superconducting shim coils for a 7 T animal MRI system, focusing on the impact of safety margin, size, and adjustability of the second‐order shim coils on the MRI system's optimization.
Methods
The study employs a nonlinear optimization method to determine the parameters of the shim coils, considering the size of the coil, the level of undesired harmonics, and the whole number approximation of the turns in each coil. The study also conducts a thorough robustness analysis, examining the effects of coil winding accuracy, former processing accuracy, and assembly angle accuracy on the harmonic intensity of each coil.
Results
The optimization design results for the 7 T MRI system's shim coils show that the magnetic field changes are less than 0.5 %. After second‐order shimming and the harmonic coupling an, the low‐order harmonics are minimized, resulting in an improved magnetic field peak‐to‐peak uniformity from 254.47 to 8.970 ppm.
Conclusion
The study successfully demonstrates the creation of a set of second‐order shim coils for a 7 T animal MRI system through numerical optimization. The design outputs provide essential technological support for the development of a human whole‐body 7 T MRI system, ensuring high‐quality imaging at the neuronal level. The project also highlights the importance of considering manufacturing and assembly flaws in the shim coil design process to achieve effective shimming in practical engineering scenarios.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Plasma Physics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, China.
- Science Island Branch, Graduate School of USTC, University of Science and Technology of China, Hefei, China.
- Institute of Energy, Hefei Comprehensive National Science Center, Hefei, China.
- School of Mechanical Engineering, Anhui University of Science & Technology, Huainan, China.
Förderung
- Institute of Energy, Hefei Comprehensive National Science Center
- Collaborative Innovation Project of Colleges and Universities of Anhui Province
Zitieren
APA
Lin, Z., Zhang, Z., Li, J., Gao, Z., Chu, Z., Liu, Y., Zhang, P., Wu, L., & Zhou, C. (2025). Harmonic analysis and optimization for closed‐loop superconducting shim coils of 7 T MRI magnet. Medical Physics. https://doi.org/10.1002/mp.17641
Vancouver
Lin Z, Zhang Z, Li J, Gao Z, Chu Z, Liu Y, et al. Harmonic analysis and optimization for closed‐loop superconducting shim coils of 7 T MRI magnet. Medical Physics. 2025. doi:10.1002/mp.17641
BibTeX
@article{Lin2025,
author = {Lin, Zijie and Zhang, Zhan and Li, Jiaxin and Gao, Zhaoyao and Chu, Zhenyu and Liu, Yongsuo and Zhang, Panfeng and Wu, Leping and Zhou, Chao},
title = {Harmonic analysis and optimization for closed‐loop superconducting shim coils of 7 T MRI magnet},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17641}
}
EUVIMED