An Optimised Multishot EPI Protocol for Ex Vivo Mouse Brain Diffusion MRI With Gd‐DOTA Contrast Enhancement
NMR in Biomedicine · 2026-04 · 39 (4)
DOI: 10.1002/nbm.70270 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift NMR in Biomedicine, online erschienen am 14.03.2026. Verfasst von Elise Cosenza und 6 weiteren Autor:innen. Beteiligte Einrichtung: Groupe d'Imagerie Neurofonctionnelle, Institut Des Maladies Neurodégénératives (GIN-IMN), UMR5293, CNRS, CEA, Université de Bordeaux, Bordeaux, Franc…. Hauptthemen laut MEDLINE: Contrast Media, Brain, Organometallic Compounds, Heterocyclic Compounds. Die Arbeit nennt 59 Literaturangaben. Gefördert von: University of Bordeaux's France 2030 program / RRI IMPACT, Agence Nationale de la Recherche.
Abstract (Originalsprache)
Diffusion‐weighted magnetic resonance imaging (dMRI) in small animal models is often limited by long acquisition times, low signal‐to‐noise ratio (SNR) and magnetic susceptibility artefacts. We developed and optimised a 3D spin‐echo multishot EPI (SE‐msEPI) protocol combined with Gd‐DOTA administration to enable efficient, high‐quality ex vivo diffusion imaging of the mouse brain. Systematic testing demonstrated that 12 EPI shots provided the optimal trade‐off between artefact reduction, SNR and acquisition duration. Gd‐DOTA incubation at 3 mM for 7 days allowed the TR to be reduced from 1500 to 400 ms without compromising diffusion metrics, yielding more than threefold SNR increase. This optimisation enabled whole‐brain acquisitions at 100‐μm isotropic resolution with 60 directions ( b = 3000 s/mm 2 ) in ~11 h, a fourfold acceleration compared to standard long‐TR protocols. Fractional anisotropy values in Gd‐DOTA–enhanced acquisitions closely matched those of the long‐TR reference, while low‐SNR short‐TR baseline and postwashout acquisitions exhibited systematic FA underestimation. Residual error analysis and tractography performance metrics further demonstrated the improved image quality and anatomically plausible reconstruction of white matter pathways under Gd‐DOTA–enhanced conditions. Beyond these technical gains, the optimised protocol provides a reproducible framework for high‐throughput ex vivo diffusion MRI, enabling connectomics and microstructural studies in mouse models with greater efficiency, increased reuse of tissue sample and reduced animal use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Groupe d'Imagerie Neurofonctionnelle, Institut Des Maladies Neurodégénératives (GIN-IMN), UMR5293, CNRS, CEA, Université de Bordeaux, Bordeaux, France.
- Centre de Résonance Magnétique des Systèmes Biologiques (CRMSB), UMR5536 CNRS, Université de Bordeaux, Bordeaux, France.
- IRP/LIA OpTeam, CNRS Biologie et Université de Bordeaux, France-Université de Sherbrooke, Canada, Bordeaux, France.
- Sherbrooke Connectivity Imaging Laboratory (SCIL), Department of Computer Science, Faculty of Sciences, Université de Sherbrooke, Sherbrooke, Quebec, Canada.
- Plateforme d'Imagerie Biomédicale (pIBIO), UAR3767 CNRS, Université de Bordeaux, Bordeaux, France.
ORCID-Profile: Elise Cosenza, Laurent Petit
Förderung
- University of Bordeaux's France 2030 program / RRI IMPACT
- Agence Nationale de la Recherche
Zitieren
APA
Cosenza, E., Corbin, N., Ribot, E. J., Boré, A., Descoteaux, M., Miraux, S., & Petit, L. (2026). An Optimised Multishot EPI Protocol for Ex Vivo Mouse Brain Diffusion MRI With Gd‐DOTA Contrast Enhancement. NMR in Biomedicine, 39(4). https://doi.org/10.1002/nbm.70270
Vancouver
Cosenza E, Corbin N, Ribot EJ, Boré A, Descoteaux M, Miraux S, et al. An Optimised Multishot EPI Protocol for Ex Vivo Mouse Brain Diffusion MRI With Gd‐DOTA Contrast Enhancement. NMR in Biomedicine. 2026;39(4). doi:10.1002/nbm.70270
BibTeX
@article{Cosenza2026,
author = {Cosenza, Elise and Corbin, Nadège and Ribot, Emeline J and Boré, Arnaud and Descoteaux, Maxime and Miraux, Sylvain and Petit, Laurent},
title = {An Optimised Multishot EPI Protocol for Ex Vivo Mouse Brain Diffusion MRI With Gd‐DOTA Contrast Enhancement},
journal = {NMR in Biomedicine},
year = {2026},
volume = {39},
number = {4},
doi = {10.1002/nbm.70270}
}
EUVIMED