Elliptical Halbach magnet and gradient modules for low‐field portable magnetic resonance imaging

Fernando Galve iD, Eduardo Pallás, Teresa Guallart-Naval, Pablo García-Cristóbal, Pablo Martínez, José M Algarín, Jose Borreguero, Rubén Bosch, Francisco Juan-Lloris, José M Benlloch, Joseba Alonso iD

NMR in Biomedicine · 2024

DOI: 10.1002/nbm.5258 ↗

Kurz gefasst

Fachartikel in der Zeitschrift NMR in Biomedicine, 2024. Verfasst von Fernando Galve und 10 weiteren Autor:innen. Beteiligte Einrichtung: MRILab, Institute for Molecular Imaging and Instrumentation (i3M), Spanish National Research Council (CSIC) and Universitat Politècnica de València (…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE European Innovation Council, Ministerio de Ciencia e Innovación, Agència Valenciana de la Innovació.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aims to develop methods to design the complete magnetic system for a truly portable MRI scanner for neurological and musculoskeletal (MSK) applications, optimized for field homogeneity, field of view (FoV), and gradient performance compared to existing low‐weight configurations. We explore optimal elliptic‐bore Halbach configurations based on discrete arrays of permanent magnets. In this way, we seek to improve the field homogeneity and remove constraints to the extent of the gradient coils typical of Halbach magnets. Specifically, we have optimized a tightly packed distribution of magnetic Nd 2 Fe 14 B cubes with differential evolution algorithms and a second array of shimming magnets with interior point and differential evolution methods. We have also designed and constructed an elliptical set of gradient coils that extend over the whole magnet length, maximizing the distance between the lobe centers. These are optimized with a target field method minimizing a cost function that considers also heat dissipation. We have employed the new toolbox to build the main magnet and gradient modules for a portable MRI scanner designed for point‐of‐care and residential use. The elliptical Halbach bore has semi‐axes of 10 and 14& cm, and the magnet generates a field of 87& mT homogeneous down to 5700& ppm (parts per million) in a 20‐cm diameter FoV; it weighs 216& kg and has a width of 65& cm and a height of 72& cm. Gradient efficiencies go up to around 0.8& mT/m/A, for a maximum of 12& mT/m within 0.5& ms with 15& A and 15& V amplifier. The distance between lobes is 28& cm, significantly increased with respect to other Halbach‐based scanners. Heat dissipation is around 25& W at maximum power, and gradient deviations from linearity are below 20% in a 20‐cm sphere. Elliptic‐bore Halbach magnets enhance the ergonomicity and field distribution of low‐cost portable MRI scanners, while allowing for full‐length gradient support to increase the FoV. This geometry can be potentially adapted for a prospective low‐cost whole‐body technology.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingMagnetsEquipment DesignPhantoms, ImagingAlgorithmsHumans

Stichwörter der Autor:innen

Halbach magnetselliptical geometrylow‐field MRIoptimization toolbox

Autor:innen und Institutionen

  • MRILab, Institute for Molecular Imaging and Instrumentation (i3M), Spanish National Research Council (CSIC) and Universitat Politècnica de València (UPV), Valencia, Spain.
  • PhysioMRI Tech S.L., Valencia, Spain.
  • Tesoro Imaging S.L., Valencia, Spain.

ORCID-Profile: Fernando Galve, Joseba Alonso

Förderung

  • HORIZON EUROPE European Innovation Council
  • Ministerio de Ciencia e Innovación
  • Agència Valenciana de la Innovació
  • European Innovation Council
  • Ministerio de Ciencia e Innovación of Spain
  • Generalitat Valenciana
  • Agencia Valenciana de la Innovación

Zitieren

APA

Galve, F., Pallás, E., Guallart-Naval, T., García-Cristóbal, P., Martínez, P., Algarín, J. M., Borreguero, J., Bosch, R., Juan-Lloris, F., Benlloch, J. M., & Alonso, J. (2024). Elliptical Halbach magnet and gradient modules for low‐field portable magnetic resonance imaging. NMR in Biomedicine. https://doi.org/10.1002/nbm.5258

Vancouver

Galve F, Pallás E, Guallart-Naval T, García-Cristóbal P, Martínez P, Algarín JM, et al. Elliptical Halbach magnet and gradient modules for low‐field portable magnetic resonance imaging. NMR in Biomedicine. 2024. doi:10.1002/nbm.5258

BibTeX
@article{Galve2024,
  author = {Galve, Fernando and Pallás, Eduardo and Guallart-Naval, Teresa and García-Cristóbal, Pablo and Martínez, Pablo and Algarín, José M and Borreguero, Jose and Bosch, Rubén and Juan-Lloris, Francisco and Benlloch, José M and Alonso, Joseba},
  title = {Elliptical Halbach magnet and gradient modules for low‐field portable magnetic resonance imaging},
  journal = {NMR in Biomedicine},
  year = {2024},
  doi = {10.1002/nbm.5258}
}
Kostenlos starten →