Signal‐to‐noise ratio versus field strength for small surface coils

Rolf Pohmann iD, Nikolai I Avdievich iD, Klaus Scheffler

NMR in Biomedicine · 2024-10 · 37 (10)

DOI: 10.1002/nbm.5168 ↗

Kurz gefasst

Fachartikel in der Zeitschrift NMR in Biomedicine, online erschienen am 08.05.2024. Verfasst von Rolf Pohmann und 2 weiteren Autor:innen. Beteiligte Einrichtung: Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany. Hauptthemen laut MEDLINE: Signal-To-Noise Ratio, Magnetic Resonance Imaging. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: European Research Council, Max-Planck-Society, Max-Planck-Gesellschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The increasing signal‐to‐noise ratio (SNR) is the main reason to use ultrahigh field MRI. Here, we investigate the dependence of the SNR on the magnetic field strength, especially for small animal applications, where small surface coils are used and coil noise cannot be ignored. Measurements were performed at five field strengths from 3 to 14.1 T, using 2.2‐cm surface coils with an identical coil design for transmit and receive on two water samples with and without salt. SNR was measured in a series of spoiled gradient echo images with varying flip angle and corrected for saturation based on a series of flip angle and T 1 measurements. Furthermore, the noise figure of the receive chain was determined and eliminated to remove instrument dependence. Finally, the coil sensitivity was determined based on the principle of reciprocity to obtain a measure for ultimate SNR. Before coil sensitivity correction, the SNR increase in nonconductive samples is highly supralinear with B 0 1.6–2.7 , depending on distance to the coil, while in the conductive sample, the growth is smaller, being around linear close to the surface coil and increasing up to a B 0 2.0 dependence when moving away from the coil. After sensitivity correction, the SNR increase is independent of loading with B 0 2.1 . This study confirms the supralinear increase of SNR with increasing field strengths. Compared with most human measurements with larger coil sizes, smaller surface coils, as mainly used in animal studies, have a higher contribution of coil noise and thus a different behavior of SNR at high fields.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Signal-To-Noise RatioMagnetic Resonance ImagingMagnetic FieldsEquipment Design

Stichwörter der Autor:innen

field strengthsignal‐to‐noise ratiosmall animal imagingsurface coilsultrahigh field

Autor:innen und Institutionen

  • Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Department of Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany.

ORCID-Profile: Rolf Pohmann, Nikolai I Avdievich

Förderung

  • European Research Council
  • Max-Planck-Society
  • Max-Planck-Gesellschaft

Zitieren

APA

Pohmann, R., Avdievich, N. I., & Scheffler, K. (2024). Signal‐to‐noise ratio versus field strength for small surface coils. NMR in Biomedicine, 37(10). https://doi.org/10.1002/nbm.5168

Vancouver

Pohmann R, Avdievich NI, Scheffler K. Signal‐to‐noise ratio versus field strength for small surface coils. NMR in Biomedicine. 2024;37(10). doi:10.1002/nbm.5168

BibTeX
@article{Pohmann2024,
  author = {Pohmann, Rolf and Avdievich, Nikolai I and Scheffler, Klaus},
  title = {Signal‐to‐noise ratio versus field strength for small surface coils},
  journal = {NMR in Biomedicine},
  year = {2024},
  volume = {37},
  number = {10},
  doi = {10.1002/nbm.5168}
}
Kostenlos starten →