EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Towards an integrated radiofrequency safety concept for implant carriers in MRI based on sensor‐equipped implants and parallel transmission

Johannes Petzold iD, Sebastian Schmitter iD, Berk Silemek iD, Lukas Winter iD, Oliver Speck iD, Bernd Ittermann iD, Frank Seifert iD

NMR in Biomedicine · 2023

DOI: 10.1002/nbm.4900 ↗

Kurz gefasst

Fachartikel in der Zeitschrift NMR in Biomedicine, 2023. Verfasst von Johannes Petzold und 6 weiteren Autor:innen. Beteiligte Einrichtung: Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany. Hauptthemen laut MEDLINE: Hot Temperature, Radio Waves. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 131 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Metrology Programme for Innovation and Research, European Partnership on Metrology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To protect implant carriers in MRI from excessive radiofrequency (RF) heating it has previously been suggested to assess that hazard via sensors on the implant. Other work recommended parallel transmission (pTx) to actively mitigate implant‐related heating. Here, both ideas are integrated into one comprehensive safety concept where native pTx safety (without implant) is ensured by state‐of‐the‐art field simulations and the implant‐specific hazard is quantified in situ using physical sensors. The concept is demonstrated by electromagnetic simulations performed on a human voxel model with a simplified spinal‐cord implant in an eight‐channel pTx body coil at . To integrate implant and native safety, the sensor signal must be calibrated in terms of an established safety metric (e.g., specific absorption rate [SAR]). Virtual experiments show that ‐field and implant‐current sensors are well suited for this purpose, while temperature sensors require some caution, and probes are inadequate. Based on an implant sensor matrix , constructed in situ from sensor readings, and precomputed native SAR limits, a vector space of safe RF excitations is determined where both global (native) and local (implant‐related) safety requirements are satisfied. Within this safe‐excitation subspace, the solution with the best image quality in terms of magnitude and homogeneity is then found by a straightforward optimization algorithm. In the investigated example, the optimized pTx shim provides a 3‐fold higher magnitude compared with circularly polarized excitation for a maximum implant‐related temperature increase . To date, sensor‐equipped implants interfaced to a pTx scanner exist as demonstrator items in research labs, but commercial devices are not yet within sight. This paper aims to demonstrate the significant benefits of such an approach and how this could impact implant‐related RF safety in MRI. Today, the responsibility for safe implant scanning lies with the implant manufacturer and the MRI operator; within the sensor concept, the MRI manufacturer would assume much of the operator's current responsibility.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hot TemperatureRadio WavesHumansComputer SimulationPhantoms, ImagingMagnetic Resonance Imaging

Stichwörter der Autor:innen

RF heatingactive implantable medical devicesimplant safetyparallel transmissionsimulationvirtual sensor

Autor:innen und Institutionen

  • Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.
  • Biomedical Magnetic Resonance, Otto von Guericke University Magdeburg, Magdeburg, Germany.

ORCID-Profile: Johannes Petzold, Sebastian Schmitter, Berk Silemek, Lukas Winter, Oliver Speck, Bernd Ittermann, Frank Seifert

Förderung

  • European Metrology Programme for Innovation and Research
  • European Partnership on Metrology

Zitieren

APA

Petzold, J., Schmitter, S., Silemek, B., Winter, L., Speck, O., Ittermann, B., & Seifert, F. (2023). Towards an integrated radiofrequency safety concept for implant carriers in MRI based on sensor‐equipped implants and parallel transmission. NMR in Biomedicine. https://doi.org/10.1002/nbm.4900

Vancouver

Petzold J, Schmitter S, Silemek B, Winter L, Speck O, Ittermann B, et al. Towards an integrated radiofrequency safety concept for implant carriers in MRI based on sensor‐equipped implants and parallel transmission. NMR in Biomedicine. 2023. doi:10.1002/nbm.4900

BibTeX
@article{Petzold2023,
  author = {Petzold, Johannes and Schmitter, Sebastian and Silemek, Berk and Winter, Lukas and Speck, Oliver and Ittermann, Bernd and Seifert, Frank},
  title = {Towards an integrated radiofrequency safety concept for implant carriers in MRI based on sensor‐equipped implants and parallel transmission},
  journal = {NMR in Biomedicine},
  year = {2023},
  doi = {10.1002/nbm.4900}
}