Numerical and experimental study of RF‐induced heating of passive implantable medical devices at 5T MRI
Magnetic Resonance in Medicine · 2025-12 · 94 (6) · 2632-2642
DOI: 10.1002/mrm.70024 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 11.08.2025. Verfasst von Mir Khadiza Akter und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Radio Waves, Prostheses and Implants. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 46 Literaturangaben. Gefördert von: Division of Industrial Innovation and Partnerships, Directorate for Technology, Innovation and Partnerships.
Abstract (Originalsprache)
Purpose
To assess the RF‐induced heating of orthopedic implants in a 5T whole‐body MRI system through electromagnetic simulations and experimental validation, with the goal of ensuring patient safety in ultra‐high field (UHF) MRI.
Methods
Numerical and experimental studies were conducted to evaluate RF‐induced heating in five titanium screws (4–12 cm) inside a 60‐cm wide 5T whole‐body MRI scanner using the standard ASTM phantom. The temperature rise over 15 min was determined through full‐wave electromagnetic simulations and direct measurements. The Finite Difference Time Domain (FDTD) method was used to quantify the 1 g mass‐averaged specific absorption rate (pSAR 1g ) in 10 clinically relevant plate‐and‐screw configurations implanted in the Duke and Ella human body models at three anatomical locations: humerus, femur, and tibia.
Results
In the phantom study, the 6 cm screw exhibited the highest SAR and temperature rise, demonstrating a resonance effect at 5T. However, in human body models, the worst‐case implant lengths shifted to 7–11 cm, highlighting the influence of tissue heterogeneity on resonance conditions. SAR values were also affected by the implant's position within the RF coil. The strong agreement between simulations and measurements validates the computational approach.
Conclusion
This study systematically evaluates RF‐induced heating in orthopedic implants within the newly approved 5T whole‐body MRI, demonstrating that implant length, positioning, and surrounding media significantly impact heating risks. The findings highlight the necessity for updated MRI safety guidelines at UHF strengths, as implant safety conditions at 5T systems may differ from those at 1.5 and 3T MRI systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA.
ORCID-Profile: Jianfeng Zheng, Ji Chen
Förderung
- Division of Industrial Innovation and Partnerships
- Directorate for Technology, Innovation and Partnerships
Zitieren
APA
Akter, M. K., Shen, A., Islam, M. Z., Zhang, L., Zheng, J., & Chen, J. (2025). Numerical and experimental study of RF‐induced heating of passive implantable medical devices at 5T MRI. Magnetic Resonance in Medicine, 94(6), 2632-2642. https://doi.org/10.1002/mrm.70024
Vancouver
Akter MK, Shen A, Islam MZ, Zhang L, Zheng J, Chen J. Numerical and experimental study of RF‐induced heating of passive implantable medical devices at 5T MRI. Magnetic Resonance in Medicine. 2025;94(6):2632-2642. doi:10.1002/mrm.70024
BibTeX
@article{Akter2025,
author = {Akter, Mir Khadiza and Shen, Ao and Islam, Md Zahidul and Zhang, Lingfei and Zheng, Jianfeng and Chen, Ji},
title = {Numerical and experimental study of RF‐induced heating of passive implantable medical devices at 5T MRI},
journal = {Magnetic Resonance in Medicine},
year = {2025},
volume = {94},
number = {6},
pages = {2632--2642},
doi = {10.1002/mrm.70024}
}