Liver T1 , T2 , T2 *, and PDFF Mapping at 0. 55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction
Magnetic Resonance in Medicine · 15.11.2026
DOI: 10.1002/mrm.70667 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 15.11.2026. Verfasst von Tom Griesler und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA. Die Arbeit nennt 64 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, Siemens Healthineers, National Institutes of Health.
Abstract (Originalsprache)
Purpose
Quantitative imaging on mid‐field (0.55 T) MRI systems can suffer from poor precision due to low SNR. This work presents an MRF pulse sequence optimized with Cramer‐Rao lower bound (CRLB) and combined with a multi‐echo deep image prior (DIP) reconstruction for co‐registered T1/T2/T2*/PDFF mapping in the liver at 0.55 T within a single breath‐hold.
Methods
A 20 s 2D multi‐echo rosette MRF breath‐hold acquisition was implemented on a 0.55 T wide‐bore system. The sequence design was optimized via pseudorandom search using a CRLB‐based objective for T1/T2 encoding and compared with reference designs. Data were reconstructed using (i) an iterative low‐rank (LR) reconstruction and (ii) a multi‐echo DIP jointly modeling TR‐ and echo‐domain dynamics including B0 effects. Water/fat separation preceded dictionary matching for T1/T2; curve fitting was used to calculate T2* and PDFF was derived from M0 with noise‐bias correction. The method was evaluated in numerical simulations, NIST and fat‐fraction phantom scans, and in vivo acquisitions in 10 healthy subjects. A linear mixed‐effects model was used to evaluate the effects of sequence design and reconstruction method on mapping accuracy and precision.
Results
Optimized sequences achieved lower CRLB cost than reference designs and, in simulations, CRLB cost was correlated with T1/T2 ROI variability ( R 2 = 0.76 LR; R 2 = 0.83 DIP). Phantom experiments showed agreement with reference values. In vivo, both sequence optimization and DIP reconstruction significantly reduced the standard deviation of all four parameters in liver ROIs.
Conclusion
CRLB‐guided acquisition and multi‐echo DIP reconstruction enable accurate and precise liver MRF T1/T2/T2*/PDFF mapping at 0.55 T in a single breath‐hold.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Biostatistics, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Internal Medicine I, University Hospital Wuerzburg, Wuerzburg, Germany.
ORCID-Profile: Tom Griesler, Jesse Hamilton, Evan Cummings, Gastao Cruz
Förderung
- NIH HHS
- Siemens Healthineers
- National Institutes of Health
Zitieren
APA
Griesler, T., Hamilton, J., Cummings, E., Richardson, J., Whiteman, A., Gram, M., Cruz, G., Hussain, H., & Seiberlich, N. (2026). Liver T1 , T2 , T2 *, and PDFF Mapping at 0. 55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70667
Vancouver
Griesler T, Hamilton J, Cummings E, Richardson J, Whiteman A, Gram M, et al. Liver T1 , T2 , T2 *, and PDFF Mapping at 0. 55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70667
BibTeX
@article{Griesler2026,
author = {Griesler, Tom and Hamilton, Jesse and Cummings, Evan and Richardson, Jacob and Whiteman, Andrew and Gram, Maximilian and Cruz, Gastao and Hussain, Hero and Seiberlich, Nicole},
title = {Liver T1 , T2 , T2 *, and PDFF Mapping at 0. 55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70667}
}
EUVIMED