Whole‐Brain Diffusion MRI at 7T With Universal Multiband GRAPE pTx Pulses
Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 146-158
DOI: 10.1002/mrm.70588 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 14.09.2026. Verfasst von Marten Laufenberg und 5 weiteren Autor:innen. Beteiligte Einrichtung: German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union Horizon 2020 Research and Innovation program, German Federal Ministry of Education and Research (BMBF).
Abstract (Originalsprache)
Purpose
To improve signal strength and homogeneity of whole‐brain diffusion‐weighted imaging (DWI) at 7T.
Methods
Universal multiband excitation and refocusing pulses were optimized with the gradient ascent pulse engineering (GRAPE) algorithm across a database of and maps obtained from 20 subjects. Slice‐selective multiband pulses were optimized in a two‐step optimization for transverse and sagittal slice orientations. The optimized pTx pulses were compared to standard circular polarized (CP) pulses in five subjects using a multiband diffusion‐weighted EPI sequence at 1.5 mm isotropic resolution and in one subject at 1.05 mm resolution. The evaluation included Bloch simulations, flip angle mapping, SNR analysis, and diffusion model fitting (DTI and NODDI).
Results
Both GRAPE pulse sets with transverse and sagittal slice orientation improved signal homogeneity and SNR compared to CP pulses across all subjects and head sizes, with particularly high gains in the cerebellum. Sagittal GRAPE pulses yielded the highest mean signal and lowest standard deviation across the brain, while transverse GRAPE pulses provided better cerebellar signal at the cost of signal loss in off‐resonant frontal and temporal regions of large heads. The improved signal translated into more reliable DTI and NODDI fitting, especially in the cerebellum.
Conclusion
Universal multiband GRAPE pTx pulses substantially improve signal homogeneity and SNR in whole‐brain DWI at 7T, enabling more reliable diffusion microstructure mapping, particularly in the cerebellum.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
- Université Paris-Saclay, Commissariat à l'Energie Atomique, CNRS, NeuroSpin, BAOBAB, Gif sur Yvette, France.
- Department of Parkinson's Disease, Sleep and Movement Disorders, Center of Neurology, University Hospital Bonn, Bonn, Germany.
- Department of Neuroradiology, University Hospital Bonn, Bonn, Germany.
- Department of Physics & Astronomy, University of Bonn, Bonn, Germany.
ORCID-Profile: Marten Laufenberg, Daniel Löwen, Nicolas Boulant, Tony Stöcker
Förderung
- European Union Horizon 2020 Research and Innovation program
- German Federal Ministry of Education and Research (BMBF)
Zitieren
APA
Laufenberg, M., Löwen, D., Ferreira, M., Boulant, N., Faber, J., & Stöcker, T. (2027). Whole‐Brain Diffusion MRI at 7T With Universal Multiband GRAPE pTx Pulses. Magnetic Resonance in Medicine, 97(1), 146-158. https://doi.org/10.1002/mrm.70588
Vancouver
Laufenberg M, Löwen D, Ferreira M, Boulant N, Faber J, Stöcker T. Whole‐Brain Diffusion MRI at 7T With Universal Multiband GRAPE pTx Pulses. Magnetic Resonance in Medicine. 2027;97(1):146-158. doi:10.1002/mrm.70588
BibTeX
@article{Laufenberg2027,
author = {Laufenberg, Marten and Löwen, Daniel and Ferreira, Mónica and Boulant, Nicolas and Faber, Jennifer and Stöcker, Tony},
title = {Whole‐Brain Diffusion MRI at 7T With Universal Multiband GRAPE pTx Pulses},
journal = {Magnetic Resonance in Medicine},
year = {2027},
volume = {97},
number = {1},
pages = {146--158},
doi = {10.1002/mrm.70588}
}