Optimizing IMPULSED Acquisition Protocols for Clinical 3T Scanners Through Bayesian Experimental Design
Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 100-113
DOI: 10.1002/mrm.70579 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 14.09.2026. Verfasst von Yan Dai und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, VROC Seed Funding 2024, UT Southwestern Medical Center, National Cancer Institute.
Abstract (Originalsprache)
Purpose
To optimize diffusion MRI acquisition protocols for Imaging Microstructural Parameters Using Limited Spectrally Edited Diffusion (IMPULSED) model at clinical 3T scanner using Bayesian experimental design, enabling accurate cellular‐scale parameter estimation under realistic scan time and scanner hardware constraints.
Methods
Expected Information Gain (EIG) was used as the optimization objective to maximize the information content of acquired measurements for IMPULSED model fitting. Bayesian optimization with Gaussian process surrogates efficiently searched the high‐dimensional acquisition parameter space, including pulse types (PGSE, OGSEn1, and OGSEn2), diffusion times, and b values. Optimized protocols were systematically evaluated against a heuristically designed baseline protocol through simulation studies assessing classification accuracy and parameter estimation performance across SNR levels of 5–40. Robustness to optimization assumptions was examined by varying prior distributions and assumed SNR. In vivo validation was performed using canine tumor data acquired at 3T.
Results
The optimized protocol eliminated OGSEn2 acquisitions, concentrated measurements at high b values (maximum 2000 s/mm 2 for PGSE, 1220 s/mm 2 for OGSEn1), employing concurrently optimized diffusion timing. Compared to the baseline protocol, the optimized design achieved superior classification accuracy for distinguishing cell populations and reduced parameter estimation error across biologically relevant parameter ranges at various SNRs. Performance advantages were consistent across diverse optimization scenarios, demonstrating robustness to prior knowledge and noise assumptions. In vivo parameter maps showed substantially improved quality and smoothness.
Conclusion
Bayesian optimization substantially improves IMPULSED acquisition design for clinical 3T scanners. This principled, algorithm‐agnostic framework enables accurate diffusion MRI cytometry under clinical constraints, with potential applications to tumor characterization and treatment monitoring.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, Maryland, USA.
ORCID-Profile: Yan Dai, Arely Perez Rodriguez
Förderung
- NCI NIH HHS
- VROC Seed Funding 2024, UT Southwestern Medical Center
- National Cancer Institute
Zitieren
APA
Dai, Y., Jia, X., Aguilera, T., Jiang, K., Perez Rodriguez, A., Vanhaezebrouck, I., & Deng, J. (2027). Optimizing IMPULSED Acquisition Protocols for Clinical 3T Scanners Through Bayesian Experimental Design. Magnetic Resonance in Medicine, 97(1), 100-113. https://doi.org/10.1002/mrm.70579
Vancouver
Dai Y, Jia X, Aguilera T, Jiang K, Perez Rodriguez A, Vanhaezebrouck I, et al. Optimizing IMPULSED Acquisition Protocols for Clinical 3T Scanners Through Bayesian Experimental Design. Magnetic Resonance in Medicine. 2027;97(1):100-113. doi:10.1002/mrm.70579
BibTeX
@article{Dai2027,
author = {Dai, Yan and Jia, Xun and Aguilera, Todd and Jiang, Kai and Perez Rodriguez, Arely and Vanhaezebrouck, Isabelle and Deng, Jie},
title = {Optimizing IMPULSED Acquisition Protocols for Clinical 3T Scanners Through Bayesian Experimental Design},
journal = {Magnetic Resonance in Medicine},
year = {2027},
volume = {97},
number = {1},
pages = {100--113},
doi = {10.1002/mrm.70579}
}
EUVIMED