EUVIMEDEUVIMED entdecken
Crossref

A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion‐Prepared MRI Acquisition Using a Multi‐Coil AC / DC Shim‐Array

Congyu Liao iD, Jason P Stockmann iD, Zhitao Li, Zhixing Wang iD, Mengze Gao iD, Lincoln Craven-Brightman, Monika Sliwiak, Charles Biggs, Jack A Glad, Jiazheng Zhou, Yurui Qian, Zheng Zhong und 9 weitere

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70294 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Congyu Liao und 20 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Brain. Die Arbeit nennt 78 Literaturangaben. Gefördert von: NIH HHS, NICHD NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

We developed a dynamic B 0 shimming approach using a 46‐channel AC/DC shim array to correct phase errors caused by eddy currents from diffusion‐encoding gradients in diffusion‐prepared MRI, enabling high b value imaging without the SNR loss from the use of a magnitude stabilizer.

Methods

A 46‐channel AC/DC shim array and corresponding amplifier system were built. Spin echo prescans with and without diffusion preparation were then used to rapidly measure eddy current–induced phase differences. These phase maps were used as targets in an optimization framework to compute compensatory shim currents for multi‐shot 3D diffusion‐prepared acquisitions.

Results

The proposed method allows flexible use of the AC/DC shim array to correct undesirable eddy current effects in diffusion‐prepared MRI. Phantom and in vivo experiments demonstrate whole‐brain, cardiac‐gated, multi‐shot 3D diffusion‐prepared imaging without the use of magnitude stabilizers. The approach enables preservation of full SNR while achieving reliable diffusion encoding at b values up to 2000 s/mm 2 .

Conclusions

This work demonstrates a new strategy for applying an AC/DC shim array to compensate for eddy current–induced phase errors in diffusion‐prepared MRI. By eliminating the need for a magnitude stabilizer, it enables efficient high‐quality diffusion imaging with full signal sensitivity retained.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diffusion Magnetic Resonance ImagingBrainPhantoms, ImagingImage EnhancementAlgorithmsEquipment DesignReproducibility of ResultsHumansArtifactsSignal-To-Noise RatioImaging, Three-DimensionalSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

B0 shim array coilconcomitant fieldsdiffusion MRIeddy currents correctionlocal field control

Autor:innen und Institutionen

  • Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA.
  • Department of Radiology, Stanford University, Stanford, California, USA.
  • Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, Arkansas, USA.
  • Department of Electrical Engineering, Stanford University, Stanford, California, USA.
  • Stanford Center for Cognitive and Neurobiological Imaging, Stanford University, Stanford, California, USA.

ORCID-Profile: Congyu Liao, Jason P Stockmann, Zhixing Wang, Mengze Gao, Hua Wu, Bernhard Gruber, Xiaozhi Cao

Förderung

  • NIH HHS
  • NICHD NIH HHS
  • National Institutes of Health

Zitieren

APA

Liao, C., Stockmann, J. P., Li, Z., Wang, Z., Gao, M., Craven-Brightman, L., Sliwiak, M., Biggs, C., Glad, J. A., Zhou, J., Qian, Y., Zhong, Z., Wang, N., Wu, H., Grafendorfer, T., Robb, F., Gruber, B., Mareyam, A., Kerr, A. B., …, & Setsompop, K. (2026). A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion‐Prepared MRI Acquisition Using a Multi‐Coil AC / DC Shim‐Array. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70294

Vancouver

Liao C, Stockmann JP, Li Z, Wang Z, Gao M, Craven-Brightman L, et al. A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion‐Prepared MRI Acquisition Using a Multi‐Coil AC / DC Shim‐Array. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70294

BibTeX
@article{Liao2026,
  author = {Liao, Congyu and Stockmann, Jason P and Li, Zhitao and Wang, Zhixing and Gao, Mengze and Craven-Brightman, Lincoln and Sliwiak, Monika and Biggs, Charles and Glad, Jack A and Zhou, Jiazheng and Qian, Yurui and Zhong, Zheng and Wang, Nan and Wu, Hua and Grafendorfer, Thomas and Robb, Fraser and Gruber, Bernhard and Mareyam, Azma and Kerr, Adam B and Cao, Xiaozhi and Setsompop, Kawin},
  title = {A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion‐Prepared MRI Acquisition Using a Multi‐Coil AC / DC Shim‐Array},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70294}
}
Kostenlos starten →