EUVIMEDObjevte EUVIMED
Open access · CC BY 4.0

Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering

Philipp Braun iD, Jonathan Stelter, Stefan Ruschke iD, Johannes M Peeters, Holger Eggers, Kilian Weiss, Daniela Junker, Dimitrios C Karampinos

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70474 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2026. Written by Philipp Braun and 7 co-authors. Institution: Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM University Hospital, Technical University of Munich (TUM), …. Major MEDLINE topics: Adipose Tissue, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Cited 1 times so far (Crossref); 63 references. The full text is freely available (open access, CC BY 4.0). Funded by: Deutsche Forschungsgemeinschaft.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

To develop a phase‐corrected time‐interleaved multi‐echo gradient echo Cartesian acquisition with spiral profile ordering (TIMGRECASPR) for abdominal large‐FOV proton density fat fraction (PDFF) mapping at 3 T , demonstrating its sampling flexibility and inherent self‐gating capabilities at high isotropic resolutions.

Methods

A methodological framework was developed to correct eddy current‐ and concomitant gradient‐related phase errors in a TIMGRECASPR acquisition. Sequence parameters were optimized with an XCAT simulation. Validation was conducted in phantoms and in vivo thigh and abdominal imaging. In five healthy volunteers, liver PDFF and T2 *, alongside pancreatic PDFF, were compared to a clinical Cartesian breath‐hold vendor technique. Liver PDFF homogeneity was evaluated visually and across five different axial subvolumes. Mean absolute PDFF bias and within‐subject standard deviation (wSD) across repeated scans were quantified. Pancreatic delineation was quantified by line profiles across the tissue boundary.

Results

Short shot lengths were required for a sufficient temporal resolution of the respiratory motion self‐navigation. Without phase corrections, data showed severe peripheral PDFF bias, especially at higher spatial resolutions. In a water phantom, PDFF bias ranged from −5% to 5% over 400 mm in readout direction, successfully mitigated by the proposed phase corrections. Liver PDFF maps exhibited improved homogeneity and accurate quantification results, verified by the breath‐hold reference scan (0.6% PDFF bias, wSD = 0.11%). Pancreatic boundaries were clearly visualized and provided accurate pancreatic PDFF maps (1.1% PDFF bias, wSD = 0.37%).

Conclusion

This work demonstrates the feasibility of a free‐breathing monopolar time‐interleaved gradient echo sequence based on a Cartesian spiral acquisition at 3 T when accounting for phase errors.

Subjects

MeSH terms (MEDLINE)

Adipose TissueMagnetic Resonance ImagingImage Processing, Computer-AssistedHumansPhantoms, ImagingLiverAlgorithmsAbdomenBreath HoldingMaleRespirationPancreasAdultFemaleReproducibility of ResultsThighHealthy VolunteersProtons

Author keywords

Cartesian spiral profile ordering (CASPR)free‐breathingisotropic resolutionphase errorsproton density fat fraction (PDFF)time‐interleaved multi‐echo gradient echo

Substances

Protons

Authors and institutions

  • Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM University Hospital, Technical University of Munich (TUM), Munich, Germany.
  • Philips Healthcare, Best, the Netherlands.
  • Philips GmbH Innovative Technologies, Hamburg, Germany.
  • Philips GmbH Market DACH, Hamburg, Germany.
  • Laboratory of Magnetic Resonance Imaging Systems and Methods, Ecole Polytechnique Federale de Lausanne (EPFL), Lausanne, Switzerland.
  • CIBM Center for Biomedical Imaging (CIBM), Lausanne, Switzerland.

ORCID profiles: Philipp Braun, Stefan Ruschke

Funding

  • Deutsche Forschungsgemeinschaft

Cite

APA

Braun, P., Stelter, J., Ruschke, S., Peeters, J. M., Eggers, H., Weiss, K., Junker, D., & Karampinos, D. C. (2026). Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70474

Vancouver

Braun P, Stelter J, Ruschke S, Peeters JM, Eggers H, Weiss K, et al. Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70474

BibTeX
@article{Braun2026,
  author = {Braun, Philipp and Stelter, Jonathan and Ruschke, Stefan and Peeters, Johannes M and Eggers, Holger and Weiss, Kilian and Junker, Daniela and Karampinos, Dimitrios C},
  title = {Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70474}
}
Start for free →