Evaluationsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Padriac Hooper und 5 weiteren Autor:innen. Beteiligte Einrichtung: ARC Training Centre for Innovation in Biomedical Imaging Technology, Brisbane, Australia. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Magnetite Nanoparticles. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 79 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Australian Research Council.
Abstract (Originalsprache)
Purpose
To measure magnetic susceptibility ( χ ) with Quantitative Susceptibility Mapping (QSM) and evaluate its repeatability using four phantom doping materials relevant to QSM applications.
Methods
A cylindrical phantom was constructed containing vials of agarose gel doped with two paramagnetic materials (ferritin, USPIO) and two diamagnetic materials (CaCl 2 , CaCO 3 ) at five concentrations each. Single orientation QSM measurements (MEDI+0) were carried out on the phantom at 3 and 7 T. We measured molar susceptibility ( χ mol ) from QSM and evaluated the precision of QSM measurements using the standard deviation of the ROI measurement (SD ROI ). We evaluated material lifespan by conducting a t ‐test of χ mol at various timepoints. Results χ mol (ppm L mmol −1 ) were measured as 1.67 ± 0.24 and 0.74 ± 0.09 (USPIO: 3 and 7 T, respectively), 10 −2 × (8.13 ± 1.35; 8.13 ± 1.19) (Ferritin: 3; 7 T), 10 −4 × (−2.68 ± 0.24; −2.71 ± 0.37) (CaCl 2 : 3; 7 T), and 10 −5 × (−9.52 ± 1.44; −9.53 ± 1.18) (CaCO 3 : 3; 7 T). We observed no significant changes in molar susceptibility for ferritin and CaCO 3 over the measured timeframes (24 and 15 months, respectively).
Conclusion
We recommend using ferritin as a paramagnetic dopant. Further research is required to identify a diamagnetic dopant with a lower electrical conductivity and a lower ratio of R 2 */ B 0 to χ .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- ARC Training Centre for Innovation in Biomedical Imaging Technology, Brisbane, Australia.
- Centre for Advanced Imaging, The University of Queensland, Brisbane, Australia.
- Siemens Healthcare Pty Ltd, Brisbane, Australia.
- School for Electrical Engineering and Computer Science, The University of Queensland, Brisbane, Australia.
- High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria.
- Christian Doppler Laboratory for MR Imaging Biomarkers, Department for Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Padriac Hooper, Jin Jin, Simon Daniel Robinson, Markus Barth
Förderung
- Australian Research Council
Zitieren
APA
Hooper, P., Jin, J., O'Brien, K., Tourell, M., Robinson, S. D., & Barth, M. (2026). Evaluation of Phantom Doping Materials in Quantitative Susceptibility Mapping. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70427
Vancouver
Hooper P, Jin J, O'Brien K, Tourell M, Robinson SD, Barth M. Evaluation of Phantom Doping Materials in Quantitative Susceptibility Mapping. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70427
BibTeX
@article{Hooper2026,
author = {Hooper, Padriac and Jin, Jin and O'Brien, Kieran and Tourell, Monique and Robinson, Simon Daniel and Barth, Markus},
title = {Evaluation of Phantom Doping Materials in Quantitative Susceptibility Mapping},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70427}
}
EUVIMED