EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry

Moritz Gutt iD, Dominik Horstmann iD, Othmar Belker iD, Simon Schröer iD, Daniel Düx, Frank Wacker, Marcel Gutberlet, Bennet Hensen

Magnetic Resonance in Medicine · 2026-09 · 96 (3) · 1274-1282

DOI: 10.1002/mrm.70392 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2026-04-22. Written by Moritz Gutt and 7 co-authors. Institution: Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. Major MEDLINE topics: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Liver, Thermometry. 39 references. The full text is freely available (open access, CC BY 4.0). Funded by: Bundesministerium für Bildung und Forschung.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

To evaluate a three dimensional dual echo magnetic resonance fingerprinting sequence for simultaneous proton resonance frequency shift (PRFS) and thermometry during microwave ablation with improving robustness near gas bubble induced susceptibility changes while PRFS provides high precision elsewhere.

Methods

An ex vivo bovine liver was heated at 40 W for 8 min. A 3D stack‐of‐spiral FLASH‐MRF acquisition on a 3 T system produced simultaneous PRFS‐ and ‐based thermometry. Two fiber optic probes served as references. was calibrated to temperature using a parenchymal region in PRFS thermometry. Thermal dose was computed as CEM43 and compared with post‐cooling inversion‐recovery imaging.

Results

The –temperature slope was . In non‐heated parenchyma, temporal standard deviation was for PRFS and for . At probe 1, RMSE was for PRFS and for . At a susceptibility‐prone probe, PRFS RMSE was while RMSE was regarding only time points up to a CEM43 threshold of 240 min. Dose maps achieved Dice overlap of 67.13% for PRFS and 73.80% for .

Conclusion

A single three dimensional MRF acquisition provided complementary temperature information. PRFS delivered high precision where phase was stable. remained monotonic near susceptibility and improved dose overlap. The framework encoded and enabled in‐session calibration from PRFS.

Subjects

MeSH terms (MEDLINE)

Imaging, Three-DimensionalMagnetic Resonance ImagingLiverThermometryAnimalsCattleMicrowavesProtons

Author keywords

MR thermometrymagnetic resonance fingerprintingmicrowave ablation

Substances

Protons

Authors and institutions

  • Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Research Campus STIMULATE, Magdeburg, Germany.

ORCID profiles: Moritz Gutt, Dominik Horstmann, Othmar Belker, Simon Schröer

Funding

  • Bundesministerium für Bildung und Forschung

Cite

APA

Gutt, M., Horstmann, D., Belker, O., Schröer, S., Düx, D., Wacker, F., Gutberlet, M., & Hensen, B. (2026). A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry. Magnetic Resonance in Medicine, 96(3), 1274-1282. https://doi.org/10.1002/mrm.70392

Vancouver

Gutt M, Horstmann D, Belker O, Schröer S, Düx D, Wacker F, et al. A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry. Magnetic Resonance in Medicine. 2026;96(3):1274-1282. doi:10.1002/mrm.70392

BibTeX
@article{Gutt2026,
  author = {Gutt, Moritz and Horstmann, Dominik and Belker, Othmar and Schröer, Simon and Düx, Daniel and Wacker, Frank and Gutberlet, Marcel and Hensen, Bennet},
  title = {A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {3},
  pages = {1274--1282},
  doi = {10.1002/mrm.70392}
}
Start for free →