A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry
Magnetic Resonance in Medicine · 2026-09 · 96 (3) · 1274-1282
DOI: 10.1002/mrm.70392 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 22.04.2026. Verfasst von Moritz Gutt und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Liver, Thermometry. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Bildung und Forschung.
Abstract (Originalsprache)
Purpose
To evaluate a three dimensional dual echo magnetic resonance fingerprinting sequence for simultaneous proton resonance frequency shift (PRFS) and thermometry during microwave ablation with improving robustness near gas bubble induced susceptibility changes while PRFS provides high precision elsewhere.
Methods
An ex vivo bovine liver was heated at 40 W for 8 min. A 3D stack‐of‐spiral FLASH‐MRF acquisition on a 3 T system produced simultaneous PRFS‐ and ‐based thermometry. Two fiber optic probes served as references. was calibrated to temperature using a parenchymal region in PRFS thermometry. Thermal dose was computed as CEM43 and compared with post‐cooling inversion‐recovery imaging.
Results
The –temperature slope was . In non‐heated parenchyma, temporal standard deviation was for PRFS and for . At probe 1, RMSE was for PRFS and for . At a susceptibility‐prone probe, PRFS RMSE was while RMSE was regarding only time points up to a CEM43 threshold of 240 min. Dose maps achieved Dice overlap of 67.13% for PRFS and 73.80% for .
Conclusion
A single three dimensional MRF acquisition provided complementary temperature information. PRFS delivered high precision where phase was stable. remained monotonic near susceptibility and improved dose overlap. The framework encoded and enabled in‐session calibration from PRFS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
- Research Campus STIMULATE, Magdeburg, Germany.
ORCID-Profile: Moritz Gutt, Dominik Horstmann, Othmar Belker, Simon Schröer
Förderung
- Bundesministerium für Bildung und Forschung
Zitieren
APA
Gutt, M., Horstmann, D., Belker, O., Schröer, S., Düx, D., Wacker, F., Gutberlet, M., & Hensen, B. (2026). A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry. Magnetic Resonance in Medicine, 96(3), 1274-1282. https://doi.org/10.1002/mrm.70392
Vancouver
Gutt M, Horstmann D, Belker O, Schröer S, Düx D, Wacker F, et al. A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry. Magnetic Resonance in Medicine. 2026;96(3):1274-1282. doi:10.1002/mrm.70392
BibTeX
@article{Gutt2026,
author = {Gutt, Moritz and Horstmann, Dominik and Belker, Othmar and Schröer, Simon and Düx, Daniel and Wacker, Frank and Gutberlet, Marcel and Hensen, Bennet},
title = {A Magnetic Resonance Fingerprinting Approach for Simultaneous T1‐ and PRFS ‐Based 3D MR ‐Thermometry},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {3},
pages = {1274--1282},
doi = {10.1002/mrm.70392}
}
EUVIMED