Using U‐Nets to Predict the Effects of Head Motion on Simulated Specific Absorption Rate for Ultra‐High Field Magnetic Resonance Imaging With Parallel Transmission
Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 650-665
DOI: 10.1002/mrm.70363 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 29.03.2026. Verfasst von Katherine Anna Blanter und 4 weiteren Autor:innen. Beteiligte Einrichtung: Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, Glamorgan, UK. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Head Movements, Head. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council, [Wales Data Nation Accelerator Award] Engineering and Physical Sciences Research Council, Wellcome Trust.
Abstract (Originalsprache)
Purpose
Ultrahigh‐field MRI requires careful management of the specific absorption rate (SAR), which is subject and subject‐position dependent. Within‐scan subject motion may exacerbate local SAR exposure, necessitating large safety margins to prevent SAR underestimation, which hampers imaging performance. This study proposes a U‐Net architecture to adapt safety calculations to motion as it happens, to facilitate high‐performance scanning without compromising safety.
Methods
Electromagnetic simulations were performed for five body models at multiple positions with an 8‐channel parallel‐transmit coil. Q‐matrices were transformed into real‐valued SAR distributions—to train U‐Nets to estimate motion‐induced effects on local SAR—which were then mapped back to Q‐matrices. Separate U‐Nets were trained for different types of body motion (rightward/leftward/anterior/posterior/yaw), which were then cascaded to predict the effect of composite (off‐axis) and larger displacements on SAR. Finally, network‐estimated local SAR distributions were compared with ground truth after‐motion local SAR for realistic parallel‐transmit pulses.
Results
Subject motion had a statistically significant effect on local SAR, but network‐estimated safety models recovered a faithful representation of the ground truth after‐motion local SAR. For the investigated parallel‐transmit pulses, the proposed approach reduced the safety margin from 2.14‐fold to 1.3‐fold and ensured more than 68% of the imaging performance could be realized while a safety model that includes all simulated subject positions would have limited scanning performance to as low as 21% of the maximum.
Conclusions
The proposed position‐aware SAR calculation approach allows smaller safety margins, which has the potential to enable higher‐performance UHF‐MRI scanning without compromising safety for subjects who are unable to remain still.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, Glamorgan, UK.
- Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Ankara, Türkiye.
- Imaging Physics & Engineering Research Department and School of Biomedical Engineering and Imaging Sciences, King's College London, London, Greater London, UK.
ORCID-Profile: Katherine Anna Blanter, Alix Plumley, Shaihan Malik, Emre Kopanoglu
Förderung
- Engineering and Physical Sciences Research Council
- [Wales Data Nation Accelerator Award] Engineering and Physical Sciences Research Council
- Wellcome Trust
- Llywodraeth Cymru
Zitieren
APA
Blanter, K. A., Plumley, A., Gungor, A., Malik, S., & Kopanoglu, E. (2026). Using U‐Nets to Predict the Effects of Head Motion on Simulated Specific Absorption Rate for Ultra‐High Field Magnetic Resonance Imaging With Parallel Transmission. Magnetic Resonance in Medicine, 96(2), 650-665. https://doi.org/10.1002/mrm.70363
Vancouver
Blanter KA, Plumley A, Gungor A, Malik S, Kopanoglu E. Using U‐Nets to Predict the Effects of Head Motion on Simulated Specific Absorption Rate for Ultra‐High Field Magnetic Resonance Imaging With Parallel Transmission. Magnetic Resonance in Medicine. 2026;96(2):650-665. doi:10.1002/mrm.70363
BibTeX
@article{Blanter2026,
author = {Blanter, Katherine Anna and Plumley, Alix and Gungor, Alper and Malik, Shaihan and Kopanoglu, Emre},
title = {Using U‐Nets to Predict the Effects of Head Motion on Simulated Specific Absorption Rate for Ultra‐High Field Magnetic Resonance Imaging With Parallel Transmission},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {2},
pages = {650--665},
doi = {10.1002/mrm.70363}
}