Effects of RF ‐Induced Heating of Orthopedic Implants Across Stages of Bone Fracture Healing and Anatomical Variability: In Silico Assessment Using Computational Human Models at 1.5 T and 3.0 T MRI

Ananya Nandikanti iD, Jianfeng Zheng, Stuart A Long, Ji Chen

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70296 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Ananya Nandikanti und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Fracture Healing, Prostheses and Implants, Fractures, Bone. Die Arbeit nennt 35 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study presents an in‐silico assessment of RF‐induced heating of orthopedic implants during various stages of bone fracture healing, namely surgery, scar, and healed phases, under 1.5 T and 3.0 T MRI.

Methods

Computational human models Duke, Ella, and Fats were used to develop the stages of bone fracture healing. These models reflect clinically relevant conditions and also include assessment of the surgery model with the presence of tissue adhesive and its leakage. Two types of orthopedic implants are analyzed: (a) a tibial nail system embedded within bone, and (b) a reverse shoulder arthroplasty (RSA) implant located partially in bone and muscle.

Results

Results demonstrate variations in RF‐induced heating across stages of bone fracture healing and anatomical models, indicating that localized tissue properties, location of the implantation as well as the presence of tissue adhesive significantly influence RF‐induced heating.

Conclusion

The findings highlight the importance of (a) integrating clinically relevant in silico models to evaluate the safety of patients undergoing MRI scans which are not included in the current testing standards for passive implantable medical devices, and (b) accurate labeling of orthopedic implants for testing in an MRI environment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingFracture HealingProstheses and ImplantsFractures, BoneHumansComputer SimulationRadio WavesHot TemperatureModels, Biological

Stichwörter der Autor:innen

RF‐induced heatingSAR1g near the devicein silico assessmentorthopedic implantspassive implantable medical devicepeak SAR1gstages of bone fracture healingtemperature rise near the devicetissue adhesive

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA.

ORCID-Profile: Ananya Nandikanti

Zitieren

APA

Nandikanti, A., Zheng, J., Long, S. A., & Chen, J. (2026). Effects of RF ‐Induced Heating of Orthopedic Implants Across Stages of Bone Fracture Healing and Anatomical Variability: In Silico Assessment Using Computational Human Models at 1.5 T and 3.0 T MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70296

Vancouver

Nandikanti A, Zheng J, Long SA, Chen J. Effects of RF ‐Induced Heating of Orthopedic Implants Across Stages of Bone Fracture Healing and Anatomical Variability: In Silico Assessment Using Computational Human Models at 1.5 T and 3.0 T MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70296

BibTeX
@article{Nandikanti2026,
  author = {Nandikanti, Ananya and Zheng, Jianfeng and Long, Stuart A and Chen, Ji},
  title = {Effects of RF ‐Induced Heating of Orthopedic Implants Across Stages of Bone Fracture Healing and Anatomical Variability: In Silico Assessment Using Computational Human Models at 1.5 T and 3.0 T MRI},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70296}
}
Kostenlos starten →