Modelling Motion‐Induced Signal Corruption in Steady‐State Diffusion MRI
Magnetic Resonance in Medicine · 2026-07 · 96 (1) · 66-82
DOI: 10.1002/mrm.70315 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 27.02.2026. Verfasst von Benjamin C Tendler und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Brain, Artifacts, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Royal Academy of Engineering, NIHR Oxford Health Biomedical Research Centre, Wellcome Trust.
Abstract (Originalsprache)
Purpose
Diffusion‐weighted steady‐state free precession (DW‐SSFP) is a diffusion imaging sequence achieving high SNR efficiency. A key challenge for in vivo DW‐SSFP is the sequence's severe motion sensitivity, currently limiting investigations to low or no motion regimes. Here we establish a framework to both (1) model and (2) correct for the impact of subject motion associated with the underlying magnetisation distribution of DW‐SSFP. Theory and Methods An extended phase graphs (EPG) representation of the 1D DW‐SSFP signal was established incorporating a motion operator describing rigid body and pulsatile motion. The representation was validated using Monte Carlo simulations, and subsequently integrated into a data fitting routine for motion estimation and correction. The fitting routine was evaluated using both simulations and a voxelwise correction applied to in vivo experimental 2D low‐resolution single‐shot timeseries DW‐SSFP data acquired in the human brain in three healthy volunteers, with a tensor reconstructed from the motion‐corrected experimental DW‐SSFP data.
Results
The proposed EPG‐motion framework gives excellent agreement to complementary Monte Carlo simulations, demonstrating that diffusion coefficient estimation is robust over a range of motion and SNR regimes. Tensor estimates from the motion‐corrected experimental DW‐SSFP data give good visual agreement to complementary diffusion‐weighted spin‐echo (DW‐SE) data acquired in the same subject, considerably reducing orientation‐dependent motion‐induced biases.
Conclusion
Temporal information capturing the evolution of the DW‐SSFP signal can be used to retrospectively (1) estimate subject motion and (2) reconstruct motion‐corrected DW‐SSFP data. Open‐source software is provided, facilitating future investigations into the impact of subject‐motion on DW‐SSFP acquisitions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
ORCID-Profile: Benjamin C Tendler, Aaron T Hess
Förderung
- Royal Academy of Engineering
- NIHR Oxford Health Biomedical Research Centre
- Wellcome Trust
Zitieren
APA
Tendler, B. C., Wu, W., Miller, K. L., & Hess, A. T. (2026). Modelling Motion‐Induced Signal Corruption in Steady‐State Diffusion MRI. Magnetic Resonance in Medicine, 96(1), 66-82. https://doi.org/10.1002/mrm.70315
Vancouver
Tendler BC, Wu W, Miller KL, Hess AT. Modelling Motion‐Induced Signal Corruption in Steady‐State Diffusion MRI. Magnetic Resonance in Medicine. 2026;96(1):66-82. doi:10.1002/mrm.70315
BibTeX
@article{Tendler2026,
author = {Tendler, Benjamin C and Wu, Wenchuan and Miller, Karla L and Hess, Aaron T},
title = {Modelling Motion‐Induced Signal Corruption in Steady‐State Diffusion MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {1},
pages = {66--82},
doi = {10.1002/mrm.70315}
}
EUVIMED