Electromagnetic Noise Characterization and Suppression in Low‐Field MRI Systems
Magnetic Resonance in Medicine · 2026-05 · 95 (5) · 3000-3007
DOI: 10.1002/mrm.70235 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 16.01.2026. Verfasst von Teresa Guallart-Naval und 2 weiteren Autor:innen. Beteiligte Einrichtung: MRILab, Institute for Molecular Imaging and Instrumentation (i3M), Spanish National Research Council (CSIC) and Universitat Politècnica de València (…. Hauptthemen laut MEDLINE: Signal-To-Noise Ratio, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministerio de Ciencia e Innovación, European Innovation Council and Small and Medium-sized Enterprises Executive Agency, European Partnership on Metrology.
Abstract (Originalsprache)
Purpose
Our goal is to develop and validate a practical protocol that guides users in identifying and suppressing electromagnetic noise in low‐field MRI systems, enabling operation near the thermal noise limit.
Methods
We present a systematic, stepwise methodology that includes diagnostic measurements, hardware isolation strategies, and good practices for cabling and shielding. Each step is validated with corresponding noise measurements under increasingly complex system configurations, both unloaded and with a human subject present.
Results
Noise levels were monitored through the incremental assembly of a low‐field MRI system, revealing key sources of EMI and quantifying their impact. Final configurations achieved noise within 1.5 the theoretical thermal bound with a subject in the scanner. Image reconstructions illustrate the direct relationship between system noise and image quality.
Conclusion
The proposed protocol enables low‐field MRI systems to operate close to fundamental noise limits in realistic conditions. The framework also provides actionable guidance for the integration of additional system components, such as gradient drivers and automatic tuning networks, without compromising signal‐to‐noise ratio (SNR).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MRILab, Institute for Molecular Imaging and Instrumentation (i3M), Spanish National Research Council (CSIC) and Universitat Politècnica de València (UPV), Valencia, Spain.
ORCID-Profile: Teresa Guallart-Naval, José Miguel Algarín, Joseba Alonso
Förderung
- Ministerio de Ciencia e Innovación
- European Innovation Council and Small and Medium-sized Enterprises Executive Agency
- European Partnership on Metrology
Zitieren
APA
Guallart-Naval, T., Algarín, J. M., & Alonso, J. (2026). Electromagnetic Noise Characterization and Suppression in Low‐Field MRI Systems. Magnetic Resonance in Medicine, 95(5), 3000-3007. https://doi.org/10.1002/mrm.70235
Vancouver
Guallart-Naval T, Algarín JM, Alonso J. Electromagnetic Noise Characterization and Suppression in Low‐Field MRI Systems. Magnetic Resonance in Medicine. 2026;95(5):3000-3007. doi:10.1002/mrm.70235
BibTeX
@article{GuallartNaval2026,
author = {Guallart-Naval, Teresa and Algarín, José Miguel and Alonso, Joseba},
title = {Electromagnetic Noise Characterization and Suppression in Low‐Field MRI Systems},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {95},
number = {5},
pages = {3000--3007},
doi = {10.1002/mrm.70235}
}
EUVIMED