Assessing radiofrequency safety of active implants by measuring induced radiofrequency currents using MRI
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70084 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Chiara Hartmann und 2 weiteren Autor:innen. Beteiligte Einrichtung: U1216, Grenoble Institut Neurosciences, Univ. Grenoble Alpes, Inserm, Grenoble, 38000, France. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Radio Waves, Prostheses and Implants. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Agence Nationale de la Recherche, Univ. Grenoble Alpes, Université Grenoble Alpes.
Abstract (Originalsprache)
Purpose
During MRI in the presence of active wire‐like implants, such as deep brain stimulation leads, there is a risk of thermal lesions in tissues adjacent to implant contacts due to radiofrequency currents induced in the wire. Currently, there is no established method to evaluate the radiofrequency (RF) safety of an implant in situ, due to complex interactions between the implant and the electric field inside the patient during MRI. This article presents a method to quantify the RF current in an implant using MRI acquisitions at very low SAR. Theory and Methods To measure RF current in situ, a modified ‐mapping sequence is proposed to image the associated perturbation of the field. A forward signal model links the RF current intensity to the MRI signal and is used to fit the RF current from acquired data. Electromagnetic simulations and experiments on a homogeneous phantom are presented for simplified and real implant wires to validate the method.
Results
The presented model can correctly reconstruct RF current amplitudes from field maps obtained with detailed electromagnetic simulations, with a normalized RMS error of 4.7%. Phantom experiments show a good linearity between the square of the current measured by MRI and temperature increase (), demonstrating that the RF current measurements quantitatively represent the effective heating.
Conclusion
A method has been developed to quantify the RF current in situ from MRI signals. This method enables to predict the individual heating risk for other MRI sequences performed in the same scanning session.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- U1216, Grenoble Institut Neurosciences, Univ. Grenoble Alpes, Inserm, Grenoble, 38000, France.
- KF-EMV, BAKOM, Biel, Switzerland.
ORCID-Profile: Chiara Hartmann, Jan M Warnking
Förderung
- Agence Nationale de la Recherche
- Univ. Grenoble Alpes
- Université Grenoble Alpes
Zitieren
APA
Hartmann, C., Bouldi, M., & Warnking, J. M. (2026). Assessing radiofrequency safety of active implants by measuring induced radiofrequency currents using MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70084
Vancouver
Hartmann C, Bouldi M, Warnking JM. Assessing radiofrequency safety of active implants by measuring induced radiofrequency currents using MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70084
BibTeX
@article{Hartmann2026,
author = {Hartmann, Chiara and Bouldi, Mélina and Warnking, Jan M},
title = {Assessing radiofrequency safety of active implants by measuring induced radiofrequency currents using MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70084}
}
EUVIMED