The effect of frequency (64–498 MHz) on specific absorption rate adjacent to metallic orthopedic screws in MRI: A numerical simulation study
Medical Physics · 2024
DOI: 10.1002/mp.16902 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Paul Jacobs und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Electromagnetic Fields, Radio Waves. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIGMS NIH HHS, FDA HHS, NCATS NIH HHS.
Abstract (Originalsprache)
Background
The imaging of patients with implanted electrically‐conductive devices via magnetic resonance imaging at ultra‐high fields is hampered by uncertainties relating to the potential for inducing tissue heating adjacent to the implant due to coupling of energy from the incident electromagnetic field into the implant. Existing data in the peer‐reviewed literature of comparisons across field strengths of tissue heating and its surrogate, the specific absorption rate (SAR), is scarce and contradictory, leading to further doubts pertaining to the safety of imaging patients with such devices.
Purpose
The radiofrequency‐induced SAR adjacent to orthopedic screws of varying length and at frequencies of 64 to 498 MHz was investigated via full‐wave electromagnetic simulations, to provide an accurate comparison of SAR across MRI field strengths.
Methods
Dipole antennas were used for RF transmission to achieve a uniform electric field tangential to the screws located 120 mm above the antenna midpoints, embedded in a bone‐mimicking material. The input power to the antennas was constrained to achieve the following targets without the screw present: (i) E = 100 V/m, (ii) B 1 + = 2 μT, and (iii) global‐average‐SAR = 3.2 W/kg. Simulations were performed with a spatial resolution of 0.2 mm in the volume surrounding the screws, resulting in 76–137 MCells, noting the maximum 1 g‐averaged SAR value in each case. Simulations were repeated at 128 and 297 MHz for screws embedded in muscle tissue.
Results
The peak SAR, occurring at the resonant screw length, substantially increased as the frequency decreased when the input power to the dipole antenna was constrained to achieve constant electric field in background tissue at the screws’ locations. A similar pattern was observed when constraining input power to achieve constant B 1 + and global‐average‐SAR. The dielectric properties of the tissue in which the screws were embedded dominated the SAR comparisons between 297 and 128 MHz.
Conclusions
The study design allowed for a direct comparison to be performed of SAR across frequencies and implant lengths without the confounding effect of variable incident electric field. Lower frequencies produced substantially larger SAR values for implants approaching the resonant length for the worst‐case uniform incident electric field along the screws’ length. The data may inform risk‐benefit assessments for imaging patients with orthopedic implants at the new clinical field strength of 7 Tesla.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Förderung
- NIGMS NIH HHS
- FDA HHS
- NCATS NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Jacobs, P., & Fagan, A. J. (2024). The effect of frequency (64–498 MHz) on specific absorption rate adjacent to metallic orthopedic screws in MRI: A numerical simulation study. Medical Physics. https://doi.org/10.1002/mp.16902
Vancouver
Jacobs P, Fagan AJ. The effect of frequency (64–498 MHz) on specific absorption rate adjacent to metallic orthopedic screws in MRI: A numerical simulation study. Medical Physics. 2024. doi:10.1002/mp.16902
BibTeX
@article{Jacobs2024,
author = {Jacobs, Paul and Fagan, Andrew J},
title = {The effect of frequency (64–498 MHz) on specific absorption rate adjacent to metallic orthopedic screws in MRI: A numerical simulation study},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16902}
}
EUVIMED