Enhanced spectral response in frequency‐dependent diffusion measurements using a linear encoding model

Eric Seth Michael iD, Franciszek Hennel iD, Klaas Paul Pruessmann iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.70006 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Eric Seth Michael und 2 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Brain, Algorithms, Diffusion Magnetic Resonance Imaging, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To devise a more comprehensive quantitative representation for spectral encodings in frequency‐dependent diffusion measurements for improved estimation of D (ω). Theory and Methods Whereas a spectral diffusion measurement is typically represented by a Dirac delta function at a single attributed frequency, spectral response is represented here by the encoding power in | Q (ω)| 2 over a set of contiguous frequency intervals. Using this representation paradigm, a linear encoding model is formulated wherein diffusivity over each interval can be estimated by inverting the encoding process from a set of measurements. This strategy was validated in in vivo human brain imaging experiments evaluating D (ω) up to 50 Hz over 10‐Hz intervals using high‐performance gradients. The employed spectral encodings were selected using an accompanying framework devised to ensure robust encoding performance given the chosen frequency intervals. Additionally, simulated measurements were carried out to compare the performance in estimating D (ω) using the proposed encoding model versus using single‐frequency attribution in relation to the form of D (ω) and the width of frequency intervals.

Results

In vivo D (ω) determined using the proposed encoding strategy were found to increase with increasing frequency and could be mapped to spectral responses more spectrally selective than those characteristic of single‐frequency attribution. In turn, simulated measurements demonstrated that the linear encoding model permitted D (ω) estimation with improved accuracy, especially for more nonlinear D (ω), at the expense of reduced precision, particularly for narrower frequency intervals.

Conclusion

By devising a more holistic representation paradigm for frequency‐dependent diffusion measurements, D (ω) can be recovered with higher fidelity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrainAlgorithmsDiffusion Magnetic Resonance ImagingImage Interpretation, Computer-AssistedImage EnhancementHumansReproducibility of ResultsLinear ModelsSensitivity and SpecificityComputer SimulationSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

diffusion MRIfrequency‐dependent diffusionoscillating gradientspectral response

Autor:innen und Institutionen

  • Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.

ORCID-Profile: Eric Seth Michael, Franciszek Hennel, Klaas Paul Pruessmann

Zitieren

APA

Michael, E. S., Hennel, F., & Pruessmann, K. P. (2025). Enhanced spectral response in frequency‐dependent diffusion measurements using a linear encoding model. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70006

Vancouver

Michael ES, Hennel F, Pruessmann KP. Enhanced spectral response in frequency‐dependent diffusion measurements using a linear encoding model. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.70006

BibTeX
@article{Michael2025,
  author = {Michael, Eric Seth and Hennel, Franciszek and Pruessmann, Klaas Paul},
  title = {Enhanced spectral response in frequency‐dependent diffusion measurements using a linear encoding model},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.70006}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche