Easy‐to‐use B1+ shims for human brain imaging at 7 T
Magnetic Resonance in Medicine · 2025-11 · 94 (5) · 2010-2022
DOI: 10.1002/mrm.30617 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 24.06.2025. Verfasst von Emma J P Brouwer und 3 weiteren Autor:innen. Beteiligte Einrichtung: Spinoza Centre for Neuroimaging, Netherlands Institute for Neuroscience, Amsterdam, the Netherlands. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Amsterdam Brain and Cognition, NWO Aspasia, Toegepaste en Technische Wetenschappen, NWO.
Abstract (Originalsprache)
Purpose field inhomogeneity is a common problem in high field brain MRI ( T). Parallel‐transmit methods that adjust the field channelwise often require valuable scan time. Group‐optimized phase shims are presented to increase or attenuate the field in specific brain regions, omitting personalized calibrations and potentially enabling reduced FOV acquisitions or artifact reduction.
Methods
Channelwise maps were obtained for seven participants using an 8Tx/32Rx coil and a 7 T MRI scanner. Two regional shim settings ( shims) were calculated: one to increase the field in the cerebellum and the other to increase the field in the occipital lobe while attenuating the field in the frontal lobe. maps from five participants outside the design group were used to simulate the profiles, and seven were scanned to evaluate the implementation of the shims using maps, 3D EPI, GRE acquisitions, and a visual fMRI experiment.
Results
Both regional shim settings successfully amplified the field in the selected ROIs resulting in improved yield and increased tSNR in the 3D EPI images and fMRI experiments compared to the circularly polarized shim mode. The attenuating shim decreased in the frontal ROI, decreasing fold‐over artifacts in a reduced FOV, lowering g‐factors in accelerated scans with high undersampling factors and resulted in improved BOLD responses in the visual fMRI experiment.
Conclusion
Regional shim settings remove the need for time‐consuming, personalized measurements and calibrations. The attenuating shim allows for signal reduction within the power limits of the rf‐coil, reducing artifacts while improving the field in selected ROIs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Spinoza Centre for Neuroimaging, Netherlands Institute for Neuroscience, Amsterdam, the Netherlands.
- Computational Cognitive Neuroscience and Neuroimaging, Netherlands Institute for Neuroscience, Amsterdam, the Netherlands.
- Department of Biomedical Engineering and Physics, Amsterdam UMC, Amsterdam, the Netherlands.
- Oxford Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
ORCID-Profile: Emma J P Brouwer, Wietske van der Zwaag, Jurjen Heij, Nikos Priovoulos
Förderung
- Amsterdam Brain and Cognition
- NWO Aspasia
- Toegepaste en Technische Wetenschappen, NWO
- NWO
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Zitieren
APA
Brouwer, E. J. P., van der Zwaag, W., Heij, J., & Priovoulos, N. (2025). Easy‐to‐use B1+ shims for human brain imaging at 7 T. Magnetic Resonance in Medicine, 94(5), 2010-2022. https://doi.org/10.1002/mrm.30617
Vancouver
Brouwer EJP, van der Zwaag W, Heij J, Priovoulos N. Easy‐to‐use B1+ shims for human brain imaging at 7 T. Magnetic Resonance in Medicine. 2025;94(5):2010-2022. doi:10.1002/mrm.30617
BibTeX
@article{Brouwer2025,
author = {Brouwer, Emma J P and van der Zwaag, Wietske and Heij, Jurjen and Priovoulos, Nikos},
title = {Easy‐to‐use B1+ shims for human brain imaging at 7 T},
journal = {Magnetic Resonance in Medicine},
year = {2025},
volume = {94},
number = {5},
pages = {2010--2022},
doi = {10.1002/mrm.30617}
}
EUVIMED