Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Sheng Shen und 6 weiteren Autor:innen. Beteiligte Einrichtung: A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Breast, Magnetic Resonance Imaging, Image Interpretation, Computer-Assisted, Image Enhancement. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, Kiyomi and Ed Baird MGH Research Scholar award, NIH HHS.
Abstract (Originalsprache)
Purpose
This study aims to develop an efficient T 1 mapping approach that is free from systematic bias caused by B 1 inhomogeneity in ultralow‐field (ULF) MRI and to determine the T 1 values (without fat suppression) of breast tissues at 6.5 mT.
Methods
We describe here a method for ULF T 1 mapping using a variable flip angle approach for both mapping T 1 and B 1 . The B 1 mapping is used to correct the T 1 maps—and, unique to operation in the ULF regime, a B 1 map acquired in a phantom can be applied to the in vivo T 1 mapping data, reducing in vivo scan time. The T 1 mapping method was validated in vitro with copper sulphate phantoms at seven different concentrations. A breast phantom was scanned to test the T 1 mapping protocol. Four healthy volunteers were recruited for T 1 mapping to obtain in vivo T 1 values of breast tissues. No fat suppression was implemented in this preliminary study.
Results
The copper sulphate phantom T 1 map measurements have less than 13% deviation from the reference T 1 values, and less than 6% deviation when the reference T 1 was greater than 50 ms. Four healthy volunteers were scanned for T 1 mapping. In these in vivo breast T 1 maps, the 25th to 75th percentiles ranged from 79 to 243 ms at 6.5 mT.
Conclusion
Variable flip angle method can be used for both B 1 and T 1 mapping at ULFs, which enables efficient in vivo T 1 mapping for ULF MRI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
- Harvard Medical School, Boston, Massachusetts, USA.
- Physical Measurement Laboratory, National Institute of Standards and Technology, Boulder, Colorado, USA.
- Department of Physics, University of Colorado, Boulder, Colorado, USA.
- Image X Institute, University of Sydney, Sydney, New South Wales, Australia.
- Massachusetts General Hospital, Division of Breast Imaging, Boston, Massachusetts, USA.
ORCID-Profile: Sheng Shen, Neha Koonjoo, Kathryn E Keenan, Matthew S Rosen
Förderung
- NCI NIH HHS
- Kiyomi and Ed Baird MGH Research Scholar award
- NIH HHS
- National Institutes of Health
Zitieren
APA
Shen, S., Koonjoo, N., Ogier, S. E., Boele, T., Saksena, M. A., Keenan, K. E., & Rosen, M. S. (2025). B 1 ‐corrected breast T 1 mapping at ultralow field. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30602
Vancouver
Shen S, Koonjoo N, Ogier SE, Boele T, Saksena MA, Keenan KE, et al. B 1 ‐corrected breast T 1 mapping at ultralow field. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30602
BibTeX
@article{Shen2025,
author = {Shen, Sheng and Koonjoo, Neha and Ogier, Stephen E and Boele, Thomas and Saksena, Mansi A and Keenan, Kathryn E and Rosen, Matthew S},
title = {B 1 ‐corrected breast T 1 mapping at ultralow field},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30602}
}
EUVIMED