Myocardial perfusion mapping with intra‐arterial spin labeling: Optimization of the labeling efficiency
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30589 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Felix Spreter und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Physics, Department of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Faculty of Medicine, Universi…. Hauptthemen laut MEDLINE: Myocardial Perfusion Imaging, Magnetic Resonance Angiography. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Purpose
Intra‐arterial spin labeling (iASL) applies a labeling pulse from a transmit coil at the tip of the catheter for regional myocardial perfusion measurements during MR‐guided catheterizations. This study investigates the labeling efficiency of catheter‐mounted coils under varying conditions to enable robust iASL measurements.
Methods
The labeling efficiency was assessed by analytical calculations, numerical simulations, and in vitro measurements at 3 T. Analytical solutions were provided for various simplified field distributions, and a numerical simulation was used to calculate for different catheter coil designs using the Biot‐Savart Law. Using these fields, the RF excitation of the blood magnetization flowing around the catheter was calculated using the Bloch equation. Finally, in vitro measurements were performed in a flow phantom using loop and solenoid catheter coils. Results fields produced by the catheter coils result in a saturation of the mean magnetization when the applied pulse exceeds a threshold power. This threshold power varies between 1 and 25 mW for different coil designs, flow velocities, and geometric parameters. Conclusion iASL creates a saturation that is robust against variations in flow and coil positioning. The simulations could determine the threshold pulse power for robust iASL excitation, which needs to be adapted to applications in different vessels. Thus, iASL may provide an efficient alternative to perfusion measurements with exogenous contrast agents during MR‐guided interventions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Medical Physics, Department of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
- Department of Cardiology and Angiology, University Heart Center Freiburg-Bad Krozingen, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
ORCID-Profile: Felix Spreter, Johannes Fischer, Ali Caglar Özen, Alexander Maier, Michael Bock, Simon Reiss
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Spreter, F., Fischer, J., Özen, A. C., Maier, A., Bock, M., & Reiss, S. (2025). Myocardial perfusion mapping with intra‐arterial spin labeling: Optimization of the labeling efficiency. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30589
Vancouver
Spreter F, Fischer J, Özen AC, Maier A, Bock M, Reiss S. Myocardial perfusion mapping with intra‐arterial spin labeling: Optimization of the labeling efficiency. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30589
BibTeX
@article{Spreter2025,
author = {Spreter, Felix and Fischer, Johannes and Özen, Ali Caglar and Maier, Alexander and Bock, Michael and Reiss, Simon},
title = {Myocardial perfusion mapping with intra‐arterial spin labeling: Optimization of the labeling efficiency},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30589}
}