Basis function compression for field probe monitoring
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30471 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Paul I Dubovan und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Biophysics, Western University, London, Ontario, Canada. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Data Compression, Algorithms, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NSERC Discovery Grant (DG) program, Canada Research Chairs, NSERC Canada Graduate Scholarships - Doctoral (CGS D) Program.
Abstract (Originalsprache)
Purpose
Field monitoring using field probes allows for accurate measurement of magnetic field perturbations, such as from eddy currents, during MRI scanning. However, errors may result when the spatial variation of the fields is not well‐described by the conventionally used spherical harmonics model that has the maximum order constrained by the number of probes. The objective of this work was to develop and validate a field monitoring approach that compresses higher order spherical harmonics into a smaller set of new basis functions that can be characterized using fewer probes.
Methods
Field monitoring of acquisitions was repeated with probes in different locations. High‐order field dynamics were computed from this “calibration” data assembled from provided scans, from which compression matrices could be devised using principal component analysis. Compression matrices were then used to fit field dynamics using “compressed” basis functions with data from 16 probes, which were then used in image reconstruction. Performance was evaluated by assessing the accuracy of computed field dynamics as well as in vivo image quality. Technique generalizability was also assessed by using various acquisition and diffusion encoding strategies in the calibration.
Results
Qualitative and quantitative improvements in accuracy were observed when using the proposed fitting method compared to the conventional approach. However, compression effectiveness was influenced by the probe quantity and arrangement, and the specific acquisition data included in the calibration.
Conclusion
The ability to tailor basis functions to more compactly describe the spatial variation of field perturbations enables improved characterization of fields with rapid spatial variations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Biophysics, Western University, London, Ontario, Canada.
- Centre for Functional and Metabolic Mapping, Western University, London, Ontario, Canada.
- Lawson Health Research Institute, London, Ontario, Canada.
- Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.
- Center for Cognitive and Neurobiological Imaging, Stanford University, Stanford, California, USA.
- Department of Electrical Engineering, Stanford University, Stanford, California, USA.
ORCID-Profile: Paul I Dubovan, Gabriel Varela-Mattatall, Eric S Michael, Franciszek Hennel, Ravi S Menon, Klaas P Pruessmann
Förderung
- NSERC Discovery Grant (DG) program
- Canada Research Chairs
- NSERC Canada Graduate Scholarships - Doctoral (CGS D) Program
Zitieren
APA
Dubovan, P. I., Varela-Mattatall, G., Michael, E. S., Hennel, F., Menon, R. S., Pruessmann, K. P., Kerr, A. B., & Baron, C. A. (2025). Basis function compression for field probe monitoring. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30471
Vancouver
Dubovan PI, Varela-Mattatall G, Michael ES, Hennel F, Menon RS, Pruessmann KP, et al. Basis function compression for field probe monitoring. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30471
BibTeX
@article{Dubovan2025,
author = {Dubovan, Paul I and Varela-Mattatall, Gabriel and Michael, Eric S and Hennel, Franciszek and Menon, Ravi S and Pruessmann, Klaas P and Kerr, Adam B and Baron, Corey A},
title = {Basis function compression for field probe monitoring},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30471}
}
EUVIMED