MVP‐VSASL: measuring MicroVascular Pulsatility using velocity‐selective arterial spin labeling
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30370 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Conan Chen und 7 weiteren Autor:innen. Beteiligte Einrichtung: Center for Functional MRI, University of California San Diego, La Jolla, California, USA. Hauptthemen laut MEDLINE: Cerebrovascular Circulation, Pulsatile Flow, Magnetic Resonance Imaging, Microvessels. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Siemens Medical Solutions USA.
Abstract (Originalsprache)
Purpose
By leveraging the small‐vessel specificity of velocity‐selective arterial spin labeling (VSASL), we present a novel technique for measuring cerebral MicroVascular Pulsatility named MVP‐VSASL. Theory and Methods We present a theoretical model relating the pulsatile, cerebral blood flow‐driven VSASL signal to the microvascular pulsatility index (), a widely used metric for quantifying cardiac‐dependent fluctuations. The model describes the dependence of the of VSASL signal (denoted ) on bolus duration (an adjustable VSASL sequence parameter) and provides guidance for selecting a value of that maximizes the SNR of the measurement. The model predictions were assessed in humans using data acquired with retrospectively cardiac‐gated VSASL sequences over a broad range of values. In vivo measurements were also used to demonstrate the feasibility of whole‐brain voxel‐wise pulsatility mapping, assess intrasession repeatability of , and illustrate the potential of this method to explore an association with age.
Results
The theoretical model showed excellent agreement to the empirical data in a gray matter region of interest (average value of 0.898 0.107 across six subjects). We further showed excellent intrasession repeatability of the pulsatility measurement (, ) and the potential to characterize associations with age (, ).
Conclusion
We have introduced a novel, VSASL‐based cerebral microvascular pulsatility technique, which may facilitate investigation of cognitive disorders where damage to the microvasculature has been implicated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Center for Functional MRI, University of California San Diego, La Jolla, California, USA.
- Department of Radiology, University of California San Diego, La Jolla, California, USA.
- Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, California, USA.
- VA San Diego Healthcare System, San Diego, California, USA.
- Department of Psychiatry, University of California San Diego, La Jolla, California, USA.
- Siemens Medical Solutions, Los Angeles, California, USA.
ORCID-Profile: Conan Chen, Thomas T Liu, Divya S Bolar
Förderung
- Siemens Medical Solutions USA
Zitieren
APA
Chen, C., Barnes, R. A., Bangen, K. J., Han, F., Pfeuffer, J., Wong, E. C., Liu, T. T., & Bolar, D. S. (2025). MVP‐VSASL: measuring MicroVascular Pulsatility using velocity‐selective arterial spin labeling. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30370
Vancouver
Chen C, Barnes RA, Bangen KJ, Han F, Pfeuffer J, Wong EC, et al. MVP‐VSASL: measuring MicroVascular Pulsatility using velocity‐selective arterial spin labeling. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30370
BibTeX
@article{Chen2025,
author = {Chen, Conan and Barnes, Ryan A and Bangen, Katherine J and Han, Fei and Pfeuffer, Josef and Wong, Eric C and Liu, Thomas T and Bolar, Divya S},
title = {MVP‐VSASL: measuring MicroVascular Pulsatility using velocity‐selective arterial spin labeling},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30370}
}
EUVIMED