Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI

Luca Zilberti iD, Cristina Curreli, Alessandro Arduino iD, Umberto Zanovello iD, Fabio Baruffaldi, Oriano Bottauscio iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30263 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2025. Written by Luca Zilberti and 5 co-authors. Institution: Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy. Major MEDLINE topics: Vibration, Magnetic Resonance Imaging. Cited 5 times so far (Crossref); 38 references. The full text is freely available (open access, CC BY 4.0). Funded by: European Partnership on Metrology.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

To quantify the extent of gradient‐induced vibrations, and the magnitude of motion‐induced displacement forces (“Lenz effect”), in conductive nonmagnetic orthopedic prostheses.

Methods

The investigation is carried out through numerical simulations, for a 3 T scanner. For gradient‐induced torques and vibrations, a knee and a shoulder implant are considered, at dB / dt equal to 42 T/s (rms). For motion‐induced forces associated with the Lenz effect, a knee and a hip implant are studied, considering a patient who translates on the examination couch, or walks next to it.

Results

Gradient‐induced torques may be within the same order of magnitude as the worst case gravitational torque defined in the ASTM standards. However, for all investigated cases, they result to be lower. In vacuum, the extent of the corresponding vibration reduces with frequency. At the lowest investigated frequency (270 Hz), it keeps below 25 μm. For an implant partially embedded in bone, the extent of the vibration increases with frequency. Nevertheless, the displacement is far lower than the worst case observed in vacuum (negligible in contact with the bone; ˜1 μm or less where the implant emerges from the bone). The Lenz effect induced by the motion of the patient through the stationary magnetic field produces forces on the order of a few millinewtons (i.e., at least two orders of magnitude lower than the implant weight).

Conclusion

Comparing the results with mechanical loads caused by ordinary activities of daily living, and with the levels of tolerable micromotions, a good safety margin is confirmed.

Subjects

MeSH terms (MEDLINE)

VibrationMagnetic Resonance ImagingHumansTorqueProstheses and ImplantsMotionComputer Simulation

Author keywords

Lenz effectsLorentz forcedisplacement forcedynamic torquesimplantsmechanical vibrations

Authors and institutions

  • Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy.
  • Laboratorio di Tecnologia Medica, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy.

ORCID profiles: Luca Zilberti, Alessandro Arduino, Umberto Zanovello, Oriano Bottauscio

Funding

  • European Partnership on Metrology

Cite

APA

Zilberti, L., Curreli, C., Arduino, A., Zanovello, U., Baruffaldi, F., & Bottauscio, O. (2025). Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30263

Vancouver

Zilberti L, Curreli C, Arduino A, Zanovello U, Baruffaldi F, Bottauscio O. Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30263

BibTeX
@article{Zilberti2025,
  author = {Zilberti, Luca and Curreli, Cristina and Arduino, Alessandro and Zanovello, Umberto and Baruffaldi, Fabio and Bottauscio, Oriano},
  title = {Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30263}
}
Start for free →