Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30263 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Luca Zilberti und 5 weiteren Autor:innen. Beteiligte Einrichtung: Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy. Hauptthemen laut MEDLINE: Vibration, Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Partnership on Metrology.
Abstract (Originalsprache)
Purpose
To quantify the extent of gradient‐induced vibrations, and the magnitude of motion‐induced displacement forces (“Lenz effect”), in conductive nonmagnetic orthopedic prostheses.
Methods
The investigation is carried out through numerical simulations, for a 3 T scanner. For gradient‐induced torques and vibrations, a knee and a shoulder implant are considered, at dB / dt equal to 42 T/s (rms). For motion‐induced forces associated with the Lenz effect, a knee and a hip implant are studied, considering a patient who translates on the examination couch, or walks next to it.
Results
Gradient‐induced torques may be within the same order of magnitude as the worst case gravitational torque defined in the ASTM standards. However, for all investigated cases, they result to be lower. In vacuum, the extent of the corresponding vibration reduces with frequency. At the lowest investigated frequency (270 Hz), it keeps below 25 μm. For an implant partially embedded in bone, the extent of the vibration increases with frequency. Nevertheless, the displacement is far lower than the worst case observed in vacuum (negligible in contact with the bone; ˜1 μm or less where the implant emerges from the bone). The Lenz effect induced by the motion of the patient through the stationary magnetic field produces forces on the order of a few millinewtons (i.e., at least two orders of magnitude lower than the implant weight).
Conclusion
Comparing the results with mechanical loads caused by ordinary activities of daily living, and with the levels of tolerable micromotions, a good safety margin is confirmed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy.
- Laboratorio di Tecnologia Medica, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy.
ORCID-Profile: Luca Zilberti, Alessandro Arduino, Umberto Zanovello, Oriano Bottauscio
Förderung
- European Partnership on Metrology
Zitieren
APA
Zilberti, L., Curreli, C., Arduino, A., Zanovello, U., Baruffaldi, F., & Bottauscio, O. (2025). Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30263
Vancouver
Zilberti L, Curreli C, Arduino A, Zanovello U, Baruffaldi F, Bottauscio O. Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30263
BibTeX
@article{Zilberti2025,
author = {Zilberti, Luca and Curreli, Cristina and Arduino, Alessandro and Zanovello, Umberto and Baruffaldi, Fabio and Bottauscio, Oriano},
title = {Gradient‐induced vibrations and motion‐induced Lenz effects on conductive nonmagnetic orthopedic implants in MRI},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30263}
}
EUVIMED