Transitioning from 0.5 to 0.9 mT : Protecting against inadvertent activation of magnet mode in active implants
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30153 ↗At a glance
Übersichtsarbeit in Magnetic Resonance in Medicine, 2024. Written by Michael C Steckner and 2 co-authors. Institution: MKS Consulting, Ohio, Beachwood, USA. Major MEDLINE topics: Magnetic Resonance Imaging. Cited 6 times so far (Crossref); 41 references. The full text is freely available (open access, CC BY 4.0). Funded by: Wellcome/EPSRC Centre for Medical Engineering, National Institute for Health Research (NIHR) Clinical Research Facility, National Institute for Health and Care Research (NIHR) Biomedical Research Centre.
Abstract (original language)
The “5 gauss line” is a phrase that is likely to be familiar to everyone working with MRI, but what is its significance, how was it defined, and what changes are currently in progress? This review explores the history of 5 gauss (0.5 mT) as a threshold for protecting against inadvertently putting cardiac pacemakers, implantable cardioverter defibrillators, and other active implantable medical devices into a “magnet mode.” Additionally, it describes the background to the recent change of this threshold to 9 gauss (0.9 mT) in the International Standard IEC 60601‐2‐33 edition 4.0 that defines basic safety requirements for MRI. Practical implications of this change and some ongoing and emerging issues are also discussed.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- MKS Consulting, Ohio, Beachwood, USA.
- IEC SC62B/MT40 IEC 60601-2-33, Geneva, Switzerland.
- ISO/TC 150/SC 6/JWG 2 (IEC/SC62B), Geneva, Switzerland.
- Medicines and Healthcare Products Regulatory Agency, London, UK.
- Guy's and St Thomas' NHS Foundation Trust, London, UK.
- King's College London, London, UK.
ORCID profiles: Michael C Steckner, Geoff Charles-Edwards
Funding
- Wellcome/EPSRC Centre for Medical Engineering
- National Institute for Health Research (NIHR) Clinical Research Facility
- National Institute for Health and Care Research (NIHR) Biomedical Research Centre
- Royal Marsden Cancer Charity
- Wellcome EPSRC Centre for Medical Engineering
- National Institute for Health and Care Research
Cite
APA
Steckner, M. C., Grainger, D., & Charles-Edwards, G. (2024). Transitioning from 0.5 to 0.9 mT : Protecting against inadvertent activation of magnet mode in active implants. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30153
Vancouver
Steckner MC, Grainger D, Charles-Edwards G. Transitioning from 0.5 to 0.9 mT : Protecting against inadvertent activation of magnet mode in active implants. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30153
BibTeX
@article{Steckner2024,
author = {Steckner, Michael C and Grainger, David and Charles-Edwards, Geoff},
title = {Transitioning from 0.5 to 0.9 mT : Protecting against inadvertent activation of magnet mode in active implants},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30153}
}
EUVIMED