Sodium triple quantum MR signal extraction using a single‐pulse sequence with single quantum time efficiency
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30107 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2024. Written by Simon Reichert and 4 co-authors. Institution: Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany. Major MEDLINE topics: Magnetic Resonance Imaging, Sodium, Phantoms, Imaging, Algorithms. Cited 2 times so far (Crossref); 74 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: State of Florida, German Research Foundation, National High Magnetic Field Laboratory.
Abstract (original language)
Purpose
Sodium triple quantum (TQ) signal has been shown to be a valuable biomarker for cell viability. Despite its clinical potential, application of Sodium TQ signal is hindered by complex pulse sequences with long scan times. This study proposes a method to approximate the TQ signal using a single excitation pulse without phase cycling.
Methods
The proposed method is based on a single excitation pulse and a comparison of the free induction decay (FID) with the integral of the FID combined with a shifting reconstruction window. The TQ signal is calculated from this FID only. As a proof of concept, the method was also combined with a multi‐echo UTE imaging sequence on a 9.4 T preclinical MRI scanner for the possibility of fast TQ MRI.
Results
The extracted Sodium TQ signals of single‐pulse and spin echo FIDs were in close agreement with theory and TQ measurement by traditional three‐pulse sequence (TQ time proportional phase increment [TQTPPI)]. For 2%, 4%, and 6% agar samples, the absolute deviations of the maximum TQ signals between SE and theoretical (time proportional phase increment TQTPPI) TQ signals were less than 1.2% (2.4%), and relative deviations were less than 4.6% (6.8%). The impact of multi‐compartment systems and noise on the accuracy of the TQ signal was small for simulated data. The systematic error was
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
- Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
- Cooperative Core Facility Animal Scanner ZI, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
- National High Magnetic Field Laboratory, Florida State University, Tallahassee, Florida, USA.
ORCID profiles: Simon Reichert, Victor Schepkin, Frank G Zöllner
Funding
- State of Florida
- German Research Foundation
- National High Magnetic Field Laboratory
- National Science Foundation
- Deutsche Forschungsgemeinschaft
Cite
APA
Reichert, S., Schepkin, V., Kleimaier, D., Zöllner, F. G., & Schad, L. R. (2024). Sodium triple quantum MR signal extraction using a single‐pulse sequence with single quantum time efficiency. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30107
Vancouver
Reichert S, Schepkin V, Kleimaier D, Zöllner FG, Schad LR. Sodium triple quantum MR signal extraction using a single‐pulse sequence with single quantum time efficiency. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30107
BibTeX
@article{Reichert2024,
author = {Reichert, Simon and Schepkin, Victor and Kleimaier, Dennis and Zöllner, Frank G and Schad, Lothar R},
title = {Sodium triple quantum MR signal extraction using a single‐pulse sequence with single quantum time efficiency},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30107}
}
EUVIMED