Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 05.03.2024. Verfasst von Eric Seth Michael und 2 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Brain, Motion, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Purpose
To evaluate the utility of up to second‐order motion‐compensated diffusion encoding in multi‐shot human brain acquisitions.
Methods
Experiments were performed with high‐performance gradients using three forms of diffusion encoding motion‐compensated through different orders: conventional zeroth‐order–compensated pulsed gradients (PG), first‐order–compensated gradients (MC1), and second‐order–compensated gradients (MC2). Single‐shot acquisitions were conducted to correlate the order of motion compensation with resultant phase variability. Then, multi‐shot acquisitions were performed at varying interleaving factors. Multi‐shot images were reconstructed using three levels of shot‐to‐shot phase correction: no correction, channel‐wise phase correction based on FID navigation, and correction based on explicit phase mapping (MUSE).
Results
In single‐shot acquisitions, MC2 diffusion encoding most effectively suppressed phase variability and sensitivity to brain pulsation, yielding residual variations of about 10° and of low spatial order. Consequently, multi‐shot MC2 images were largely satisfactory without phase correction and consistently improved with the navigator correction, which yielded repeatable high‐quality images; contrarily, PG and MC1 images were inadequately corrected using the navigator approach. With respect to MUSE reconstructions, the MC2 navigator‐corrected images were in close agreement for a standard interleaving factor and considerably more reliable for higher interleaving factors, for which MUSE images were corrupted. Finally, owing to the advanced gradient hardware, the relative SNR penalty of motion‐compensated diffusion sensitization was substantially more tolerable than that faced previously.
Conclusion
Second‐order motion‐compensated diffusion encoding mitigates and simplifies shot‐to‐shot phase variability in the human brain, rendering the multi‐shot acquisition strategy an effective means to circumvent limitations of retrospective phase correction methods.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.
ORCID-Profile: Eric Seth Michael, Franciszek Hennel, Klaas Paul Pruessmann
Zitieren
APA
Michael, E. S., Hennel, F., & Pruessmann, K. P. (2024). Motion‐compensated diffusion encoding in multi‐shot human brain acquisitions: Insights using high‐performance gradients. Magnetic Resonance in Medicine, 92(2), 556-572. https://doi.org/10.1002/mrm.30069
Vancouver
Michael ES, Hennel F, Pruessmann KP. Motion‐compensated diffusion encoding in multi‐shot human brain acquisitions: Insights using high‐performance gradients. Magnetic Resonance in Medicine. 2024;92(2):556-572. doi:10.1002/mrm.30069
BibTeX
@article{Michael2024,
author = {Michael, Eric Seth and Hennel, Franciszek and Pruessmann, Klaas Paul},
title = {Motion‐compensated diffusion encoding in multi‐shot human brain acquisitions: Insights using high‐performance gradients},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {92},
number = {2},
pages = {556--572},
doi = {10.1002/mrm.30069}
}
EUVIMED