Quantifying 3D MR fingerprinting ( 3D‐MRF ) reproducibility across subjects, sessions, and scanners automatically using MNI atlases
Magnetic Resonance in Medicine · 2024-05 · 91 (5) · 2074-2088
DOI: 10.1002/mrm.29983 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 09.01.2024. Verfasst von Andrew Dupuis und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NCI NIH HHS, NICHD NIH HHS.
Abstract (Originalsprache)
Purpose
Quantitative MRI techniques such as MR fingerprinting (MRF) promise more objective and comparable measurements of tissue properties at the point‐of‐care than weighted imaging. However, few direct cross‐modal comparisons of MRF's repeatability and reproducibility versus weighted acquisitions have been performed. This work proposes a novel fully automated pipeline for quantitatively comparing cross‐modal imaging performance in vivo via atlas‐based sampling.
Methods
We acquire whole‐brain 3D‐MRF, turbo spin echo, and MPRAGE sequences three times each on two scanners across 10 subjects, for a total of 60 multimodal datasets. The proposed automated registration and analysis pipeline uses linear and nonlinear registration to align all qualitative and quantitative DICOM stacks to Montreal Neurological Institute (MNI) 152 space, then samples each dataset's native space through transformation inversion to compare performance within atlas regions across subjects, scanners, and repetitions.
Results
Voxel values within MRF‐derived maps were found to be more repeatable (σ T1 = 1.90, σ T2 = 3.20) across sessions than vendor‐reconstructed MPRAGE (σ T1w = 6.04) or turbo spin echo (σ T2w = 5.66) images. Additionally, MRF was found to be more reproducible across scanners (σ T1 = 2.21, σ T2 = 3.89) than either qualitative modality (σ T1w = 7.84, σ T2w = 7.76). Notably, differences between repeatability and reproducibility of in vivo MRF were insignificant, unlike the weighted images.
Conclusion
MRF data from many sessions and scanners can potentially be treated as a single dataset for harmonized analysis or longitudinal comparisons without the additional regularization steps needed for qualitative modalities.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Radiology, University Hospitals, Cleveland, Ohio, USA.
- Microsoft Research, Redmond, Washington, USA.
- Siemens Medical Solutions USA, Inc, Chicago, Illinois, USA.
ORCID-Profile: Andrew Dupuis, Yong Chen, Michael Hansen, Kelvin Chow, Jessie E P Sun, Chaitra Badve, Dan Ma, Mark A Griswold, Rasim Boyacioglu
Förderung
- NIBIB NIH HHS
- NCI NIH HHS
- NICHD NIH HHS
- NINDS NIH HHS
Zitieren
APA
Dupuis, A., Chen, Y., Hansen, M., Chow, K., Sun, J. E. P., Badve, C., Ma, D., Griswold, M. A., & Boyacioglu, R. (2024). Quantifying 3D MR fingerprinting ( 3D‐MRF ) reproducibility across subjects, sessions, and scanners automatically using MNI atlases. Magnetic Resonance in Medicine, 91(5), 2074-2088. https://doi.org/10.1002/mrm.29983
Vancouver
Dupuis A, Chen Y, Hansen M, Chow K, Sun JEP, Badve C, et al. Quantifying 3D MR fingerprinting ( 3D‐MRF ) reproducibility across subjects, sessions, and scanners automatically using MNI atlases. Magnetic Resonance in Medicine. 2024;91(5):2074-2088. doi:10.1002/mrm.29983
BibTeX
@article{Dupuis2024,
author = {Dupuis, Andrew and Chen, Yong and Hansen, Michael and Chow, Kelvin and Sun, Jessie E P and Badve, Chaitra and Ma, Dan and Griswold, Mark A and Boyacioglu, Rasim},
title = {Quantifying 3D MR fingerprinting ( 3D‐MRF ) reproducibility across subjects, sessions, and scanners automatically using MNI atlases},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {91},
number = {5},
pages = {2074--2088},
doi = {10.1002/mrm.29983}
}
EUVIMED