Adult brain T1 and T2 values measured at 3 T using magnetic resonance fingerprinting with phantom validation
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70289 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Kalina V Jordanova und 7 weiteren Autor:innen. Beteiligte Einrichtung: National Institute of Standards and Technology, Boulder, Colorado, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Council Postdoctoral Fellowship, NIST-PREP (Professional Research Experience Program), U.S. Department of Commerce.
Abstract (Originalsprache)
Background
For clinical implementation of quantitative MRI methods, it is necessary to understand the expected range of nominally healthy tissue properties. Additionally, to compare measurements across MRI systems or provide guidance for clinically meaningful change, these nominally healthy measurements should be completed with phantom validation of the measurement system and method.
Purpose
To measure T1 and T2 relaxation times on 20 participants using magnetic resonance fingerprinting (MRF) concurrently with phantom measurements for validation.
Methods
T1 and T2 relaxation times were measured at 3 T using MRF on 20 participants. To assess longitudinal reproducibility, a phantom was scanned in the same session as each in vivo measurement, which resulted in 11 phantom measurements over the 99‐day study duration. At each imaging session, phantom and in vivo test‐retest measurements were made. Phantom test‐retest repeatability was evaluated along with longitudinal reproducibility. Additionally, in vivo test‐retest repeatability was evaluated for brain regions of various sizes, and mean in vivo measurements for cerebrospinal fluid (CSF), gray matter (GM), and white matter (WM) are presented.
Results
Over the study duration, phantom measurements were repeatable and reproducible, and no changes were detected in the MRF measurement of T1 and T2 relaxation times on this system. For in vivo measurements, test‐retest variation for CSF, GM, and WM was 2.3%, 0.31%, 0.57% for T1 and 8.3%, 1.9%, 1.3% for T2. We observed that the test‐retest variation of T1 and T2 measurements increased as the segmented volume fell below 1 cm 3 . Mean and standard deviation of T1 and T2 were calculated over all participants for CSF, GM, and WM.
Conclusions
Phantom measurements characterized the measurement accuracy, repeatability, and reproducibility throughout the study. The MRF T1 and T2 measurements of normal brain tissues at 3 T are validated through concurrent phantom measurements, and mean values were reported for each tissue. This study provides one data point to answer the question of how often phantom quality assessment must be done in tandem with in vivo measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- National Institute of Standards and Technology, Boulder, Colorado, USA.
- Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany.
ORCID-Profile: Kathryn E Keenan
Förderung
- National Research Council Postdoctoral Fellowship
- NIST-PREP (Professional Research Experience Program)
- U.S. Department of Commerce
- NIST DOC
- National Institute of Standards and Technology
Zitieren
APA
Jordanova, K. V., Ogier, S. E., Russek, S. E., Stoffer, C. M., Stupic, K. F., Buonincontri, G., Nittka, M., & Keenan, K. E. (2025). Adult brain T1 and T2 values measured at 3 T using magnetic resonance fingerprinting with phantom validation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70289
Vancouver
Jordanova KV, Ogier SE, Russek SE, Stoffer CM, Stupic KF, Buonincontri G, et al. Adult brain T1 and T2 values measured at 3 T using magnetic resonance fingerprinting with phantom validation. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70289
BibTeX
@article{Jordanova2025,
author = {Jordanova, Kalina V and Ogier, Stephen E and Russek, Stephen E and Stoffer, Cassandra M and Stupic, Karl F and Buonincontri, Guido and Nittka, Mathias and Keenan, Kathryn E},
title = {Adult brain T1 and T2 values measured at 3 T using magnetic resonance fingerprinting with phantom validation},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70289}
}
EUVIMED