Three‐dimensional sector‐wise golden angle–improved k‐space uniformity after electrocardiogram binning

Alexander Fyrdahl iD, Amanda Ullvin, Joao G Ramos, Nicole Seiberlich, Martin Ugander, Andreas Sigfridsson

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29698 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Alexander Fyrdahl und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Physiology, Karolinska University Hospital, and Karolinska Institutet, Stockholm, Sweden. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Cine, Heart. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop and evaluate a 3D sector‐wise golden‐angle (3D‐SWIG) profile ordering scheme for cardiovascular MR cine imaging that maintains high k‐space uniformity after electrocardiogram (ECG) binning.

Method

Cardiovascular MR (CMR) was performed at 1.5 T. A balanced SSFP pulse sequence was implemented with a novel 3D‐SWIG radial ordering, where k‐space was divided into wedges, and each wedge was acquired in a separate heartbeat. The high uniformity of k‐space coverage after physiological binning can be used to perform functional imaging using a very short acquisition. The 3D‐SWIG was compared with two commonly used 3D radial trajectories for CMR (i.e., double golden angle and spiral phyllotaxis) in numerical simulations. Free‐breathing 3D‐SWIG and conventional breath‐held 2D cine were compared in patients ( n = 17) referred clinically for CMR. Quantitative comparison was performed based on left ventricular segmentation.

Results

Numerical simulations showed that 3D‐SWIG both required smaller steps between successive readouts and achieved better k‐space sampling uniformity after binning than either the double golden angle or spiral phyllotaxis trajectories. In vivo evaluation showed that measurements of left ventricular ejection fraction calculated from a 48 heart‐beat free‐breathing 3D‐SWIG acquisition were highly reproducible and agreed with breath‐held 2D‐Cartesian cine (mean ± SD difference of −3.1 ± 3.5% points).

Conclusions

The 3D‐SWIG acquisition offers a simple solution for highly improved k‐space uniformity after physiological binning. The feasibility of the 3D‐SWIG method is demonstrated in this study through whole‐heart cine imaging during free breathing with an acquisition time of less than 1 min.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalMagnetic Resonance Imaging, CineHeartHumansElectrocardiographyRespirationHeart RateStroke VolumeVentricular Function, Left

Stichwörter der Autor:innen

CMRSWIGcardiac cinefree breathinggolden anglewhole heart

Autor:innen und Institutionen

  • Department of Clinical Physiology, Karolinska University Hospital, and Karolinska Institutet, Stockholm, Sweden.
  • Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
  • The Kolling Institute, Royal North Shore Hospital, and University of Sydney, Sydney, Australia.

ORCID-Profile: Alexander Fyrdahl

Zitieren

APA

Fyrdahl, A., Ullvin, A., Ramos, J. G., Seiberlich, N., Ugander, M., & Sigfridsson, A. (2023). Three‐dimensional sector‐wise golden angle–improved k‐space uniformity after electrocardiogram binning. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29698

Vancouver

Fyrdahl A, Ullvin A, Ramos JG, Seiberlich N, Ugander M, Sigfridsson A. Three‐dimensional sector‐wise golden angle–improved k‐space uniformity after electrocardiogram binning. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29698

BibTeX
@article{Fyrdahl2023,
  author = {Fyrdahl, Alexander and Ullvin, Amanda and Ramos, Joao G and Seiberlich, Nicole and Ugander, Martin and Sigfridsson, Andreas},
  title = {Three‐dimensional sector‐wise golden angle–improved k‐space uniformity after electrocardiogram binning},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29698}
}
Kostenlos starten →