EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Performance of Free‐Running Cardiac and Respiratory Motion‐Resolved Whole‐Heart 5D MRI in Congenital Heart Disease Patients Using a Gadolinium Enhanced Fast‐Interrupted Steady‐State Sequence

Christopher W Roy iD, Eva Vogeli, Alissia Megalo, Fabienne Dirbach, Jessica A M Bastiaansen, Jérôme Yerly iD, Marco Mueller, David Rodrigues, Jean-Baptiste Ledoux iD, Juerg Schwitter iD, Milan Prša iD, Estelle Tenisch iD und 2 weitere

Journal of Magnetic Resonance Imaging · 2026

DOI: 10.1002/jmri.70402 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2026. Verfasst von Christopher W Roy und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland. Hauptthemen laut MEDLINE: Heart Defects, Congenital, Magnetic Resonance Imaging, Gadolinium, Heart. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung, Wellcome Trust.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The performance of 5D imaging in CHD has been previously demonstrated using ferumoxytol rather than gadolinium.

Purpose

To evaluate the performance of gadolinium enhanced 5D MRI relative to 2D and 3D MRI in CHD patients. Study Type Retrospective. Subjects 45 consecutive CHD patients referred for clinically indicated MRI who underwent 2D, 3D, and 5D MRI. Field Strength/Sequence 2D bSSFP CINE, 3D bSSFP, and 5D FISS at 1.5 T. Assessment Scan time (2D, 3D, and 5D), left and right ventricular (LV, RV) end diastolic and end systolic volumes (LVEDV, RVEDV, LVESV, and RVESV, respectively), ejection fraction (LVEF, RVEF) for 2D versus 5D, and image quality on a 4‐point Likert scale (3D vs. 5D). Statistical Tests Paired comparisons were made using a paired t ‐test or Wilcoxon signed rank test. Agreement between measurements was evaluated with Bland–Altman analysis including the bias and 95% limits of agreement.

Results

Scan time for 5D (6:18 [0]) was significantly shorter than 2D (7:16 [4:39]) and 3D (6:55 [1:30]) MRI. 5D measurements were strongly correlated with 2D measurements of LVEDV ( R 2 = 0.94), LVESV ( R 2 = 0.83), LVEF ( R 2 = 0.51), ( R 2 = 0.85), RVESV ( R 2 = 0.80), and RVEF ( R 2 = 0.66) but with significant biases. Sharpness of the coronary arteries were significantly higher for 3D (RCA: 36.8% ± 10.2%, LAD: 40.5% ± 10.5%) than 5D (RCA: 24.3% ± 7.5%, LAD: 31.7% ± 8.7%), but longer vessel segments were visible using 5D (LAD: 6.9 ± 2.5) than 3D (LAD: 5.3 ± 2.9 mm) acquisitions. The grades for 5D (R1: 3 [1], R2: 2.5 [0.5], R3: 3 [0.6]) were significantly higher than 3D (R1: 3 [1], R2: 2.5 [1], R3: 2 [1]). Data Conclusion Gadolinium‐enhanced 5D MRI may provide an efficient and simplified acquisition for CHD patients, with multiple quantitative parameters from 5D being not significantly different from those from reference standard 2D and 3D MRI. Evidence Level 4 Stage of Technical Efficacy 1

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Defects, CongenitalMagnetic Resonance ImagingGadoliniumHeartHumansFemaleRetrospective StudiesContrast MediaMaleImaging, Three-DimensionalAdultReproducibility of ResultsRespirationAdolescentMotionChildYoung AdultMiddle AgedMagnetic Resonance Imaging, CineImage Interpretation, Computer-AssistedStroke VolumeImage Enhancement

Stichwörter der Autor:innen

cardiac functioncongenital heart diseasecoronary magnetic resonance angiographyfree‐runningwhole‐heart

Substanzen

Contrast MediaGadolinium

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • Service of Cardiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • Department of Diagnostic, Interventional and Pediatric Radiology (DIPR), Inselspital, Bern University Hospital, Switzerland.
  • Translational Imaging Center (TIC), Swiss Institute for Translational and Entrepreneurial Medicine, Bern, Switzerland.
  • Center for Biomedical Imaging (CIBM), Lausanne, Switzerland.
  • Swiss Innovation Hub, Siemens Healthineers International AG, Lausanne, Switzerland.

ORCID-Profile: Christopher W Roy, Jérôme Yerly, Jean-Baptiste Ledoux, Juerg Schwitter, Milan Prša, Estelle Tenisch, Tobias Rutz, Matthias Stuber

Förderung

  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
  • Wellcome Trust

Zitieren

APA

Roy, C. W., Vogeli, E., Megalo, A., Dirbach, F., Bastiaansen, J. A. M., Yerly, J., Mueller, M., Rodrigues, D., Ledoux, J. B., Schwitter, J., Prša, M., Tenisch, E., Rutz, T., & Stuber, M. (2026). Performance of Free‐Running Cardiac and Respiratory Motion‐Resolved Whole‐Heart 5D MRI in Congenital Heart Disease Patients Using a Gadolinium Enhanced Fast‐Interrupted Steady‐State Sequence. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70402

Vancouver

Roy CW, Vogeli E, Megalo A, Dirbach F, Bastiaansen JAM, Yerly J, et al. Performance of Free‐Running Cardiac and Respiratory Motion‐Resolved Whole‐Heart 5D MRI in Congenital Heart Disease Patients Using a Gadolinium Enhanced Fast‐Interrupted Steady‐State Sequence. Journal of Magnetic Resonance Imaging. 2026. doi:10.1002/jmri.70402

BibTeX
@article{Roy2026,
  author = {Roy, Christopher W and Vogeli, Eva and Megalo, Alissia and Dirbach, Fabienne and Bastiaansen, Jessica A M and Yerly, Jérôme and Mueller, Marco and Rodrigues, David and Ledoux, Jean-Baptiste and Schwitter, Juerg and Prša, Milan and Tenisch, Estelle and Rutz, Tobias and Stuber, Matthias},
  title = {Performance of Free‐Running Cardiac and Respiratory Motion‐Resolved Whole‐Heart 5D MRI in Congenital Heart Disease Patients Using a Gadolinium Enhanced Fast‐Interrupted Steady‐State Sequence},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2026},
  doi = {10.1002/jmri.70402}
}
Kostenlos starten →