Bias‐corrected respiratory motion standard deviation from amplitude‐binned 4D‐MRI for radiotherapy planning

Michael Ferraro, Sirisha Tadimalla, Fidel Navarro Salazar, Eireann Cosgriff, Annette Haworth, Jonathan Sykes

Medical Physics · 2026-11 · 53 (11)

DOI: 10.1002/mp.70745 ↗

At a glance

Journal article in Medical Physics, published online on 2026-11-14. Written by Michael Ferraro and 5 co-authors. Institution: Institute of Medical Physics, University of Sydney, Sydney, NSW, Australia. Major MEDLINE topics: Magnetic Resonance Imaging, Respiration, Radiotherapy Planning, Computer-Assisted. 24 references. The full text is freely available (open access, CC BY 4.0). Funded by: Cancer Institute NSW.

SantraukaSubjectsBibliografiniai duomenysCiteRelated studies

Abstract (original language)

Background

Respiratory‐resolved magnetic resonance imaging (4D‐MRI) provides superior soft‐tissue contrast for characterizing respiratory motion in radiotherapy, but its geometric fidelity depends on the reconstruction and binning strategy. As with 4D‐CT, peak‐to‐peak amplitude derived from binned 4D‐MRI is prone to underestimation. However, as acquisitions span many respiratory cycles, the standard deviation (SD) of motion, required for probability‐based planning approaches such as mid‐ventilation, may be estimated accurately, provided binning‐induced bias is corrected.

Purpose

To develop and validate a motion SD estimator for equal‐probability amplitude‐binned 4D‐MRI data that corrects for systematic underestimation caused by intra‐bin motion variability.

Methods

4D‐MRI data were acquired on a dynamic motion phantom driven by five patient‐derived respiratory traces. Images were reconstructed using both a standard non‐uniform fast Fourier transform (NUFFT) and the compressed sensing‐based XD‐GRASP algorithm with varying numbers (2–20) of equal‐probability respiratory amplitude bins. An SD estimator based on the law of total variance was implemented to correct for intra‐bin variance. The accuracy of peak‐to‐peak amplitude and SD was evaluated against ground‐truth physical phantom motion, with performance under deformable motion tested on a digital phantom (CoMBAT).

Results

Both reconstruction methods underestimated true peak‐to‐peak motion. While the inter‐bin SD calculation also underestimated motion variability, the proposed correction substantially mitigated this bias. It yielded robust SD estimates with median residual underestimation of 0.13 mm (5 bins) and 0.11 mm (10 bins) across patient traces. In CoMBAT, the absolute bias of the total SD estimate was below 3% from five bins onward.

Conclusions

The proposed estimator provides an accurate measurement of respiratory motion SD from amplitude‐binned 4D‐MRI by accounting for intra‐bin variability. This supports the reliability of 4D‐MRI for planning strategies dependent on motion probability distributions. Additionally, the confirmed underestimation of peak‐to‐peak amplitude highlights a fundamental limitation in defining internal target volumes directly from binned 4D‐MRI.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingRespirationRadiotherapy Planning, Computer-AssistedMovementHumansPhantoms, Imaging

Author keywords

4D‐MRImid‐ventilationmotion managementradiation therapystandard deviation estimation

Authors and institutions

  • Institute of Medical Physics, University of Sydney, Sydney, NSW, Australia.
  • Image X Institute, University of Sydney, Sydney, NSW, Australia.
  • Radiation Oncology Network, Western Sydney Local Health District, Sydney, NSW, Australia.

Funding

  • Cancer Institute NSW

Cite

APA

Ferraro, M., Tadimalla, S., Navarro Salazar, F., Cosgriff, E., Haworth, A., & Sykes, J. (2026). Bias‐corrected respiratory motion standard deviation from amplitude‐binned 4D‐MRI for radiotherapy planning. Medical Physics, 53(11). https://doi.org/10.1002/mp.70745

Vancouver

Ferraro M, Tadimalla S, Navarro Salazar F, Cosgriff E, Haworth A, Sykes J. Bias‐corrected respiratory motion standard deviation from amplitude‐binned 4D‐MRI for radiotherapy planning. Medical Physics. 2026;53(11). doi:10.1002/mp.70745

BibTeX
@article{Ferraro2026,
  author = {Ferraro, Michael and Tadimalla, Sirisha and Navarro Salazar, Fidel and Cosgriff, Eireann and Haworth, Annette and Sykes, Jonathan},
  title = {Bias‐corrected respiratory motion standard deviation from amplitude‐binned 4D‐MRI for radiotherapy planning},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {11},
  doi = {10.1002/mp.70745}
}
Start for free →