Performance evaluation of a dedicated brain PET scanner with octagonal detector arrangement using a modified NEMA NU2 2018 protocol

Ekaterina Mikhaylova, Max L Ahnen, Henryk Barthel, Jan Debus, Jannis Fischer, Martin Huellner, Markus Jehl, Philipp A Kaufmann, Michael Rullmann, Bernhard Sattler, Karen G Steinhoff, Juan Streb and 4 more

Medical Physics · 2026

DOI: 10.1002/mp.70578 ↗

At a glance

Evaluationsstudie in Medical Physics, 2026. Written by Ekaterina Mikhaylova and 15 co-authors. Institution: Positrigo, Technoparkstrasse, Zurich, Switzerland. Major MEDLINE topics: Positron-Emission Tomography, Brain. 31 references. The full text is freely available in PubMed Central.

AbstraktSubjectsBibliograafilised andmedCiteRelated studies

Abstract (original language)

Background

This work presents the NEMA NU2 2018 performance evaluation of the NeuroLF Basic, a compact and fully integrated brain positron emission tomography (PET) system based on 10‐mm‐long LYSO crystals.

Purpose

The goal is to assess its quantitative performance and imaging capabilities for neuroimaging applications.

Methods

The evaluation includes measurements and simulations of spatial resolution, scatter fraction, noise equivalent count rate (NECR), accuracy of corrections, image quality, and sensitivity, following standardized protocols to ensure reliable and reproducible results. In addition, imaging of the Hoffman 3‐dimensional brain phantom was performed to visually evaluate system performance. Image quality was assessed using an ACR Esser phantom, which fits within the compact bore of the NeuroLF scanner.

Results

The average filtered‐back‐projection spatial resolution, measured at 1 and 10 cm radial offsets, was 2.51 and 3.61 mm, respectively, highlighting the system's ability to capture details essential for neuroimaging. The scatter fraction was 24.2% at the peak NECR, reflecting the system's capability to minimize scattered events. The peak NECR was 55.3 kcps at an activity concentration of 8.1 kBq/mL, demonstrating the system's efficiency in handling typical neuroimaging count rates. NeuroLF showed excellent accuracy of corrections at 3.7% and good image contrast for hot vials in a warm background in the adapted image quality procedure with the ACR Esser phantom, achieving values of 87.5%, 77%, 60%, and 35.8% for 25‐, 16‐, 12‐, and 8‐mm‐diameter vials, respectively, with a maximum background variability of 10% for the 8‐mm vial. Sensitivity, evaluated using a 70‐cm line source, was 5.85 cps/kBq at the center of the field of view and 8.08 cps/kBq at a 10 cm radial offset. Images of the Hoffman phantom further demonstrated the system's ability to produce realistic brain uptake distributions. The measured NEMA NU2 2018 results showed good agreement with Monte Carlo simulation results, validating the simulated NeuroLF model for future research and optimization.

Conclusions

These results demonstrate the NeuroLF's potential to deliver high‐quality images, good detection efficiency, and robust quantitative performance. The comprehensive NEMA NU2 2018 evaluation, supported by Hoffman phantom imaging, confirms NeuroLF's capability as a valuable tool for clinical applications in neuroimaging.

Subjects

MeSH terms (MEDLINE)

Positron-Emission TomographyBrainPhantoms, ImagingImage Processing, Computer-AssistedSignal-To-Noise Ratio

Author keywords

GATE simulationsNEMA NU2 2018brain PETperformance evaluation

Authors and institutions

  • Positrigo, Technoparkstrasse, Zurich, Switzerland.
  • Department of Nuclear Medicine, University of Leipzig Medical Center, Leipzig, Germany.
  • Department of Nuclear Medicine, University Hospital Zurich/University of Zurich, Zurich, Switzerland.

Cite

APA

Mikhaylova, E., Ahnen, M. L., Barthel, H., Debus, J., Fischer, J., Huellner, M., Jehl, M., Kaufmann, P. A., Rullmann, M., Sattler, B., Steinhoff, K. G., Streb, J., Tiepolt, S., Treyer, V., Sabri, O., & Sacco, I. (2026). Performance evaluation of a dedicated brain PET scanner with octagonal detector arrangement using a modified NEMA NU2 2018 protocol. Medical Physics. https://doi.org/10.1002/mp.70578

Vancouver

Mikhaylova E, Ahnen ML, Barthel H, Debus J, Fischer J, Huellner M, et al. Performance evaluation of a dedicated brain PET scanner with octagonal detector arrangement using a modified NEMA NU2 2018 protocol. Medical Physics. 2026. doi:10.1002/mp.70578

BibTeX
@article{Mikhaylova2026,
  author = {Mikhaylova, Ekaterina and Ahnen, Max L and Barthel, Henryk and Debus, Jan and Fischer, Jannis and Huellner, Martin and Jehl, Markus and Kaufmann, Philipp A and Rullmann, Michael and Sattler, Bernhard and Steinhoff, Karen G and Streb, Juan and Tiepolt, Solveig and Treyer, Valerie and Sabri, Osama and Sacco, Ilaria},
  title = {Performance evaluation of a dedicated brain PET scanner with octagonal detector arrangement using a modified NEMA NU2 2018 protocol},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70578}
}
Start for free →