EUVIMEDAtrasti EUVIMED
Open access

Gain correction for variable imaging geometry

Noam Weiss, Ben Ezair

Medical Physics · 2026-07 · 53 (7)

DOI: 10.1002/mp.70560 ↗

At a glance

Journal article in Medical Physics, published online on 2026-07-10. Written by Noam Weiss and 1 co-authors. Institution: Research and Technology, Medtronic, Caesarea, Israel. Major MEDLINE topics: Image Processing, Computer-Assisted. 29 references. The full text is freely available in PubMed Central.

SantraukaSubjectsBibliografiniai duomenysCiteRelated studies

Abstract (original language)

Background

Gain (flat‐field) correction is a standard preprocessing step in X‐ray imaging that compensates for detector fixed‐pattern response, electronic offset, and spatial non‐uniformity in incident fluence. Conventional correction relies on a geometry‐matched air reference acquired at the same source‐detector relative position as the clinical image. In variable‐geometry systems, changes in source‐detector distance (SDD) and relative source‐detector rotation (SDR) alter the illumination field through inverse‐square dilution and orientation‐dependent anode heel shading, making geometry‐matched reference acquisition impractical.

Purpose

To develop and evaluate a geometry‐aware gain correction method that synthesizes a geometry‐matched reference field from a compact calibration dataset and recorded acquisition geometry, enabling gain correction for variable geometry without geometry‐matched reference acquisitions.

Methods

The measured signal was modeled as a function of the dark‐current, detector response, geometry‐dependent dilution, and a source fluence map that includes heel‐effect gradients. Rigid transformations were used to map between the detector and source domains, for any acquisition geometry. Calibration consisted of a dark acquisition to estimate electronic offset, an air acquisition at a large SDD to estimate detector response, and a set of air images at a moderate SDD to construct a source fluence map. For each raw image, a geometry‐specific synthetic reference was generated, followed by standard gain correction.

Results

Synthetic reference images closely matched measured air references. In air‐only images, residual non‐uniformity after correction remained below 0.5% of the mean signal, comparable to conventional matched‐reference correction across varying SDDs and SDRs. In a chest phantom study, images corrected using synthetic and matched references demonstrated high agreement (multiscale structural similarity index = 0.995; Lin's concordance correlation coefficient = 0.999) with no visually structured residual artifacts.

Conclusions

The proposed synthetic‐reference approach enables accurate gain correction under variable imaging geometry with performance comparable to conventional matched‐reference methods, eliminating the need for geometry‐specific air reference acquisitions.

Subjects

MeSH terms (MEDLINE)

Image Processing, Computer-AssistedCalibration

Author keywords

anode heel effectflat‐field calibrationgain correctionvariable imaging geometry

Authors and institutions

  • Research and Technology, Medtronic, Caesarea, Israel.

Cite

APA

Weiss, N., & Ezair, B. (2026). Gain correction for variable imaging geometry. Medical Physics, 53(7). https://doi.org/10.1002/mp.70560

Vancouver

Weiss N, Ezair B. Gain correction for variable imaging geometry. Medical Physics. 2026;53(7). doi:10.1002/mp.70560

BibTeX
@article{Weiss2026,
  author = {Weiss, Noam and Ezair, Ben},
  title = {Gain correction for variable imaging geometry},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {7},
  doi = {10.1002/mp.70560}
}
Start for free →