EUVIMEDEUVIMED entdecken
Crossref

Deep learning‐based dose prediction for MR‐guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection

Hao-Wen Cheng, Wen-Chih Tseng, Guanghua Yan, Jonathan Li, Chihray Liu

Medical Physics · 2026-07 · 53 (7)

DOI: 10.1002/mp.70547 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 30.06.2026. Verfasst von Hao-Wen Cheng und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville, Florida, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Deep Learning, Radiotherapy Planning, Computer-Assisted, Prostatic Neoplasms. Die Arbeit nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The template‐based treatment planning workflow in the Unity magnetic resonance‐guided linear accelerator (MR‐Linac) helps facilitate the treatment planning process; however, dose optimization and calculation remain time‐consuming in complex cases. In addition, the deformable image registration (DIR) algorithm used during adaptive planning may produce inaccurate results, thereby making efficient contour editing essential for online adaptation, especially in cases involving multiple organs at risk adjacent to the target.

Purpose

This study aims to resolve key clinical challenges in treatment planning and online adaptive planning for prostate simultaneous integrated boost (SIB) treatment on the MR‐Linac using deep learning (DL)‐based dose prediction. Multiple DL‐based dose prediction models are developed and evaluated for rapid dosimetric feasibility assessment, and the top‐performing model is used to guide the development of a novel contouring strategy to streamline the adaptive planning workflow.

Methods

Six DL models were developed and evaluated for 3D dose prediction using 80 reference and adaptive plans from 20 patients previously treated with SIB intensity‐modulated radiation therapy (SIB‐IMRT) on the Unity MR‐Linac. To address DIR‐introduced contour inaccuracies during adaptation, a contour editing strategy based on a minimal manual editing margin (MMEM) around the target structures was introduced. The MMEM was determined using six margin sizes (5‐30 mm, in 5‐mm increments) evaluated by the top‐performing DL model. Twenty adaptive plans with DIR inaccuracies in bladder, rectum, and bowel bag from 12 patients were analyzed for MMEM determination, and 10 adaptive plans were further evaluated for feasibility in the Monaco treatment planning system.

Results

The dosimetric differences between model predictions and the treatment planning system‐calculated ground truth were small for most models (≤1.5 Gy or 2%), and the average Dice Similarity Coefficients across the 10–100% isodose volumes exceeded 0.9, demonstrating strong spatial agreement. Among the evaluated models, the attention‐gated (AG) U‐Net achieved the best overall performance and was used for the MMEM determination. Based on the MMEM and feasibility analyses, a 10‐mm editing margin is recommended for the rectum, with the use of the semi‐automatic segmentation tool for the bladder and no manual edits for the bowel bag.

Conclusion

The AG U‐Net enabled accurate dose prediction for rapid dosimetric feasibility assessment in prostate SIB treatment on the Unity MR‐Linac. Leveraging this model, an MMEM‐based contouring strategy was developed to guide contour modification and improve the efficiency of online adaptive planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingDeep LearningRadiotherapy Planning, Computer-AssistedProstatic NeoplasmsRadiotherapy, Image-GuidedRadiation DosageRadiotherapy, Intensity-ModulatedHumansMaleFeasibility StudiesRadiotherapy DosageTime Factors

Stichwörter der Autor:innen

MRgRTadaptive radiation therapydose prediction

Autor:innen und Institutionen

  • Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville, Florida, USA.
  • Department of Radiation Medicine, Medical University of South Carolina, Charleston, South Carolina, USA.

Zitieren

APA

Cheng, H. W., Tseng, W. C., Yan, G., Li, J., & Liu, C. (2026). Deep learning‐based dose prediction for MR‐guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection. Medical Physics, 53(7). https://doi.org/10.1002/mp.70547

Vancouver

Cheng HW, Tseng WC, Yan G, Li J, Liu C. Deep learning‐based dose prediction for MR‐guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection. Medical Physics. 2026;53(7). doi:10.1002/mp.70547

BibTeX
@article{Cheng2026,
  author = {Cheng, Hao-Wen and Tseng, Wen-Chih and Yan, Guanghua and Li, Jonathan and Liu, Chihray},
  title = {Deep learning‐based dose prediction for MR‐guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {7},
  doi = {10.1002/mp.70547}
}
Kostenlos starten →