Enhancing dose conformity in head and neck intensity‐modulated proton therapy using a novel dynamic multi‐leaf collimator strategy

Yushi Wakisaka, Yuki Tominaga, Yuya Miyasaka, Robabeh Rahimi, Keith M Furutani, Mayu Nakata, Takeo Iwai, Teiji Nishio

Medical Physics · 2026

DOI: 10.1002/mp.70355 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Yushi Wakisaka und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, Medical Co. Hakuhokai, Osaka Proton Therapy Clinic, Osaka, Japan. Hauptthemen laut MEDLINE: Proton Therapy, Radiotherapy, Intensity-Modulated, Head and Neck Neoplasms, Radiation Dosage. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Proton pencil beam scanning (PBS) enables highly conformal dose distributions; however, its lateral dose fall‐off (penumbra) can be compromised by the use of range shifters (RSs) and increased air gaps. In PBS for head and neck regions, where critical organs at risk (OARs) are frequently adjacent to the target, penumbra degradation may lead to increased OAR doses or suboptimal target coverage. The integration of a dynamic multi‐leaf collimator (dMLC), which adjusts leaf positions at each energy layer, has been shown to improve dose conformity in single‐field uniform dose (SFUD) delivery. In parallel, intensity‐modulated proton therapy (IMPT) offers enhanced dose shaping over SFUD by modulating beam intensity across multiple fields and does not require a single beam to encompass the entire target volume, providing greater flexibility in utilizing dMLC capabilities.

Purpose

This study integrates dMLC into IMPT for head and neck cancer and proposes a novel leaf positioning strategy. We evaluate the dosimetric impact of this approach and assess its potential clinical benefits in terms of target coverage and OAR sparing.

Methods

Treatment plans were retrospectively created for five patients with head and neck cancer. For each patient, IMPT plans with three beam angles were generated using three techniques: (1) uncollimated PBS, (2) dMLC cover , in which the MLC encloses the target cross‐section at each energy layer, and (3) dMLC block , in which the MLC actively blocks OARs and their distal regions. Dose‐volume histogram (DVH) metrics for the clinical target volume (CTV) and OARs were evaluated, including a total of 21 perturbed scenarios that combined ± 2 mm setup uncertainties (7 scenarios) and ± 3.5% range uncertainties (3 scenarios). The accuracy of dose calculations was validated by comparing calculated and measured lateral dose distributions at the isocenter plane in water using two‐dimensional gamma analysis with a 2%/2 mm criterion.

Results

The integration of dMLC with IMPT significantly reduced the dose to surrounding OARs while maintaining comparable target coverage and robustness relative to uncollimated PBS. Notably, dMLC block demonstrated an enhanced dose‐sparing effect than dMLC cover , particularly for OARs surrounded by the target. While maintaining comparable CTV D98% across three techniques, dMLC cover achieved the greatest reduction in the mean doses to the eyeballs and optic nerves, as well as in the D2% to the optic chiasm, brain, and brainstem in most cases. The gamma passing rate between calculated and measured doses for dMLC block exceeded 95% for all beams, confirming the accuracy of dose calculations involving complex leaf positions.

Conclusions

The combination of IMPT and dMLC provides notable dosimetric advantages, supporting its potential for clinical applications. Further validation across a broader range of cases is necessary to comprehensively assess its efficacy and safety, particularly with respect to leaf positioning accuracy and potential variations in biological effectiveness.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiotherapy, Intensity-ModulatedHead and Neck NeoplasmsRadiation DosageRadiotherapy DosageHumansRadiotherapy Planning, Computer-AssistedOrgans at Risk

Stichwörter der Autor:innen

dynamic collimationmulti leaf collimatorproton therapy

Autor:innen und Institutionen

  • Department of Radiotherapy, Medical Co. Hakuhokai, Osaka Proton Therapy Clinic, Osaka, Japan.
  • Medical Physics Laboratory, Division of Health Science, Graduate School of Medicine, The University of Osaka, Osaka, Japan.
  • Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan.
  • Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland, USA.
  • Division of Medical Physics, Department of Radiation Oncology, Mayo Clinic Florida, Jacksonville, Florida, USA.
  • Department of Carbon Ion Radiotherapy, Graduate School of Medicine, The University of Osaka, Osaka, Japan.

Förderung

  • JSPS KAKENHI

Zitieren

APA

Wakisaka, Y., Tominaga, Y., Miyasaka, Y., Rahimi, R., Furutani, K. M., Nakata, M., Iwai, T., & Nishio, T. (2026). Enhancing dose conformity in head and neck intensity‐modulated proton therapy using a novel dynamic multi‐leaf collimator strategy. Medical Physics. https://doi.org/10.1002/mp.70355

Vancouver

Wakisaka Y, Tominaga Y, Miyasaka Y, Rahimi R, Furutani KM, Nakata M, et al. Enhancing dose conformity in head and neck intensity‐modulated proton therapy using a novel dynamic multi‐leaf collimator strategy. Medical Physics. 2026. doi:10.1002/mp.70355

BibTeX
@article{Wakisaka2026,
  author = {Wakisaka, Yushi and Tominaga, Yuki and Miyasaka, Yuya and Rahimi, Robabeh and Furutani, Keith M and Nakata, Mayu and Iwai, Takeo and Nishio, Teiji},
  title = {Enhancing dose conformity in head and neck intensity‐modulated proton therapy using a novel dynamic multi‐leaf collimator strategy},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70355}
}
Kostenlos starten →