Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy

Naonori Hu, Ryo Kakino, Akinori Sasaki, Mai Nojiri, Kazuhiko Akita, Syuushi Yoshikawa, Yasushi Kohigashi, Yuki Yoshino, Satoshi Takeno, Teruhito Aihara, Takushi Takata, Hiroki Tanaka and 2 more

Medical Physics · 2026

DOI: 10.1002/mp.70319 ↗

At a glance

Journal article in Medical Physics, 2026. Written by Naonori Hu and 13 co-authors. Institution: Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan. Major MEDLINE topics: Boron Neutron Capture Therapy, Radiation Protection, Neutrons, Lithium. 20 references. The full text is freely available (open access, CC BY 4.0). Funded by: JSPS KAKENHI.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Boron neutron capture therapy (BNCT) enables selective tumor irradiation by exploiting the high‐linear energy transfer particles generated from neutron interactions with 10 B atoms. BNCT has been approved as an insurance‐covered medical treatment for recurrent head and neck cancer in Japan. Unlike photon radiotherapy, neutrons that come out of the collimator have an angular distribution. Therefore, it is necessary to keep the distance between the collimator and the patient as short as possible. However, for head and neck cancer treatments, patient anatomy often limits proximity to the collimator, creating an unwanted air gap. This ultimately increases the neutron exposure to surrounding healthy tissue.

Purpose

To develop a freely deformable LiF‐polyethylene neutron shield and assess its impact on neutron/gamma attenuation and clinical organ at‐risk sparing in head and neck BNCT.

Methods

A freely deformable neutron shielding device was constructed using polyethylene beads loaded with lithium fluoride encapsulated in a vacuum‐sealed cushion. Neutron and gamma‐ray attenuation were measured in a water phantom under clinical conditions using an accelerator‐based BNCT system (NeuCure®, Kansai BNCT Medical Center). Measurements were compared with a solid LiF‐polyethylene block and validated through Monte Carlo–based simulations in a commercial treatment planning system. Three representative head and neck cases were further simulated to assess clinical dosimetric effects.

Results

The deformable shielding device reduced the thermal neutron flux by approximately 50%, compared with 60% for the solid LiF‐polyethylene block. Simulated head and neck treatments demonstrated significant OAR dose reductions (up to 46.6% in pharyngeal mucosa D 50% ) without compromising tumor dose coverage ( D 80% ≥ 20 Gy‐eq). Treatment delivery times were minimally affected (< 2 min difference) across all plans. A 5 mm positional perturbation analysis showed ≤ 0.5 Gy‐eq variation in GTV D min and pharyngeal mucosa D 50 and D max .

Conclusions

The freely deformable LiF‐based neutron shielding device effectively attenuated stray neutron dose while maintaining target coverage in BNCT. Its adaptability and reusability make it a practical adjunct for patient‐specific dose optimization in clinical BNCT applications.

Subjects

MeSH terms (MEDLINE)

Boron Neutron Capture TherapyRadiation ProtectionNeutronsLithiumMonte Carlo MethodHumansRadiometryHead and Neck NeoplasmsPhantoms, ImagingRadiotherapy DosageRadiotherapy Planning, Computer-Assisted

Author keywords

BNCTMonte Carlo simulationneutron shielding

Substances

Lithium

Authors and institutions

  • Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan.
  • Particle Radiation Oncology Research Center, Kyoto University Institute for Integrated Radiation and Nuclear Science, Osaka, Japan.
  • BNCT Joint Clinical Institute, Osaka Medical and Pharmaceutical University, Osaka, Japan.
  • Department of Radiology, Kyoto Prefectural University of Medicine, Kyoto, Japan.
  • Department of Radiation Oncology, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.
  • Department of Otorhinolaryngology, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.

Funding

  • JSPS KAKENHI

Cite

APA

Hu, N., Kakino, R., Sasaki, A., Nojiri, M., Akita, K., Yoshikawa, S., Kohigashi, Y., Yoshino, Y., Takeno, S., Aihara, T., Takata, T., Tanaka, H., Nihei, K., & Ono, K. (2026). Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy. Medical Physics. https://doi.org/10.1002/mp.70319

Vancouver

Hu N, Kakino R, Sasaki A, Nojiri M, Akita K, Yoshikawa S, et al. Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy. Medical Physics. 2026. doi:10.1002/mp.70319

BibTeX
@article{Hu2026,
  author = {Hu, Naonori and Kakino, Ryo and Sasaki, Akinori and Nojiri, Mai and Akita, Kazuhiko and Yoshikawa, Syuushi and Kohigashi, Yasushi and Yoshino, Yuki and Takeno, Satoshi and Aihara, Teruhito and Takata, Takushi and Tanaka, Hiroki and Nihei, Keiji and Ono, Koji},
  title = {Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70319}
}
Start for free →