EUVIMEDEUVIMED entdecken
Open Access

Development of an online neutron beam monitoring method for accelerator‐based BNCT using an 8‐MeV proton‐beryllium source

Masashi Takada, Hiroaki Kumada, Masashi Oyauchi, Susumu Tanaka, Tomoya Nunomiya, Toshiyuki Ohba, Akihiko Masuda, Tetsuro Matsumoto

Medical Physics · 2026-08 · 53 (8)

DOI: 10.1002/mp.70629 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 10.08.2026. Verfasst von Masashi Takada und 7 weiteren Autor:innen. Beteiligte Einrichtung: National Defense Academy of Japan, Yokosuka, Kanagawa, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Neutrons, Beryllium, Particle Accelerators. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: University of Tsukuba, National Defense Academy of Japan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Boron neutron capture therapy (BNCT) has emerged as a promising cell‐selective clinical treatment for recurrent glioblastoma as well as head and neck cancers. In recent years, BNCT has been implemented in several hospitals using accelerator‐based neutron sources. However, direct online measurement of the incident neutron beam intensity during clinical BNCT remains a major challenge. Neutron dosimetry is typically inferred from the accumulated electric charge of the incident proton beam delivered to the beryllium neutron‐production target.

Purpose

To improve the reliability of neutron beam control, it is essential to implement an online neutron‐monitoring system that complements proton beam current measurements. Such a system enables precise real‐time regulation of the neutron dose delivered to patients. In this study, we developed a new online neutron‐beam monitoring technique capable of measuring the neutron beam intensity in real time.

Methods

Reliable monitoring of neutron intensity at the patient position was achieved by detecting fast neutrons. The fast‐neutron detector consisted of a silicon diode coupled with a high‐density polyethylene neutron converter. The spatial distribution of neutrons was evaluated using Monte Carlo simulations to determine the optimal placement of the neutron beam monitor. The proposed method was validated through gold‐activation analysis, cell irradiation experiments, and accelerator tuning.

Results

Online neutron‐beam monitoring was successfully achieved by detecting fast neutrons at the end of the vacuum beam duct, where neutrons propagate directly without traversing radiation shielding. The neutron counting rate increased immediately at the onset of proton‐beam irradiation and decreased upon its termination, exhibiting periodic fluctuations throughout the irradiation period. These fluctuations did not significantly affect the integrated neutron counts, as they were averaged over a typical BNCT irradiation time of approximately 30 min. The fast‐neutron detector produced a mean counting rate of 124.6 counts per mC of incident proton‐beam charge, with a relative standard deviation of 1.2%, and yielded a neutron‐calibration factor of per monitor unit. The neutron counting rates remained consistent even when different irradiation phantoms were placed at the patient position.

Conclusion

The online neutron‐beam monitoring technique using a fast‐neutron detector demonstrated high reproducibility, as the neutron counting rates remained stable regardless of the irradiation phantom placed downstream.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyNeutronsBerylliumParticle AcceleratorsMonte Carlo MethodHumansRadiometry

Stichwörter der Autor:innen

Monte Carlo simulationboron neutron capture therapyneutron measurementreal‐time monitoringsi diode

Substanzen

Beryllium

Autor:innen und Institutionen

  • National Defense Academy of Japan, Yokosuka, Kanagawa, Japan.
  • Proton Medical Research Center, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan.
  • Industry Business Group, Fuji Electric Co., Ltd., Hino, Tokyo, Japan.
  • NAT Co., Hitachinaka-shi, Ibaraki, Japan.
  • National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, Tsukuba, Ibaraki, Japan.

Förderung

  • University of Tsukuba
  • National Defense Academy of Japan

Zitieren

APA

Takada, M., Kumada, H., Oyauchi, M., Tanaka, S., Nunomiya, T., Ohba, T., Masuda, A., & Matsumoto, T. (2026). Development of an online neutron beam monitoring method for accelerator‐based BNCT using an 8‐MeV proton‐beryllium source. Medical Physics, 53(8). https://doi.org/10.1002/mp.70629

Vancouver

Takada M, Kumada H, Oyauchi M, Tanaka S, Nunomiya T, Ohba T, et al. Development of an online neutron beam monitoring method for accelerator‐based BNCT using an 8‐MeV proton‐beryllium source. Medical Physics. 2026;53(8). doi:10.1002/mp.70629

BibTeX
@article{Takada2026,
  author = {Takada, Masashi and Kumada, Hiroaki and Oyauchi, Masashi and Tanaka, Susumu and Nunomiya, Tomoya and Ohba, Toshiyuki and Masuda, Akihiko and Matsumoto, Tetsuro},
  title = {Development of an online neutron beam monitoring method for accelerator‐based BNCT using an 8‐MeV proton‐beryllium source},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {8},
  doi = {10.1002/mp.70629}
}
Kostenlos starten →